Hi Benjamin,
I assume from your description your instrument that runs with IconNMR is an
(the only?) instrument with a robot changer?
However, you can run IconNMR without having to have a sample changer present,
and simply configure it to use ‘Manual Sample Exchange'. This will provide
all the usual capability of IconNMR as you require, and simply needs a person
to exchange samples when one has finished. This should provide the functionality
you have described, unless I have misunderstood?
Best regards
Tim
> From: main_at_ammrl.groups.io on behalf of Atterberry, Benjamin via groups.io
> Date: Tuesday, 1 September 2026 at 20:33
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AU Program for Automated Sample Queuing in TopSpin?
Hi all,
I am the NMR Manager at Mississippi State University. Do you by any chance have
an AU program for Bruker TopSpin that runs samples one after the other, like
IconNMR does? What I'm looking for is something where you can create different
datasets in TopSpin, then queue up sample insertion, lock, automatic tuning,
topshim, getprosol, optimizing receiver gain, and zg.
We have one spectrometer that gets crowded because it has IconNMR, so the users
like it because they can queue up several 1H experiments. If we could use an AU
program to "impersonate" IconNMR on another spectrometer, it would free up this
one instrument.
Best,
Benjamin Atterberry
----------------------------------
Benjamin A. Atterberry
NMR Facility Manager
Department of Chemistry
Office: Hand Lab G052
Mississippi State University