Re: [AMMRL] AMMRL: Lab Air Pressure

From: Peter Thompson via groups.io <pmthomps=ncsu.edu_at_groups.io>
Date: Fri, 6 Sep 2024 15:16:11 -0400

Ken, we haven't tried running NEOs at lower air pressure. We have NMRs in
three locations. One used house N2 from a large exterior tank that went to
the whole building. It supplied ~85 psi. However, consumption of N2 went up
significantly and we're still in the middle of getting flow meters put into
the system to figure out where all the gas is going. In the interim, we're
using 230L N2 tanks set to higher pressure.

Our other two locations rely on rotary screw air compressors that are
utilized either solely or primarily by our facility. The downstream
pressures for these are ~110 psi, in part because we have nitrogen
generators running off this compressed air for some of our mass specs, and
they need a higher pressure than the NMRs. While being in charge of the
compressors means more work/headaches to keep them going, it also means we
have more control over when they are maintained and what extra loads do or
don't get added in. If you're going the compressor route, I definitely
second the recommendations to have a good-sized ballast tank, as well as
oil and particulate filters. And if you're running VT experiments, an extra
in-line dryer/n2 generator is a good idea as well.

Best of luck!

Peter

On Fri, Sep 6, 2024 at 1:48 PM Emily Crull via groups.io wrote:

> I haven't set one up, but all 3 of the Neo's I've worked with had
> compressors and also had filters installed to keep particulate out of the
> air line. To my understanding, it wasn't so much that we didn't have the
> pressure, but that the compressors made it so that the pressure supplied
> was more consistent.
>
>
>

--

*Peter Thompson, Ph.D.*
Lead Research Scientist, Molecular Education, Technology and Research
Innovation Center (METRIC)
Research Assistant Professor, Department of Molecular and Structural
Biochemistry
North Carolina State University
51 Polk Hall
Raleigh, NC 27695

Phone: 919.515.0836
Email: pmthomps_at_ncsu.edu
METRIC Website: https://research.ncsu.edu/metric/home
Consultation and Sample Submission: https://labs.ncsu.edu
Biochemistry Website:
https://cals.ncsu.edu/molecular-and-structural-biochemistry/

“All electronic mail messages in connection with State business which are
sent to or received by this account are subject to the NC Public Records
Law and may be disclosed to third parties.”


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1603): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1603__;!!PvDODwlR4mBZyAb0!Wv_Dr4visRJkw6CGCYHlO0dRrDqXTA=
trBJPP5u2UmM2GfdUkYD1c1ewwCfxlolnLbQy2ePCFsSSZ7SroLubi3TtEzOai$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108309346=
/7559972__;!!PvDODwlR4mBZyAb0!Wv_Dr4visRJkw6CGCYHlO0dRrDqXTAtrBJPP5u2UmM2Gf=
dUkYD1c1ewwCfxlolnLbQy2ePCFsSSZ7SroLubi3XPTItQI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 06 2024 - 12:16:25 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST