Re: [AMMRL] AMMRL: Lab Air Pressure

From: Samrat Amin <saamin1_at_mainex1.asu.edu>
Date: Fri, 06 Sep 2024 11:25:29 -0700

Hi Ken,

We have a Neo, and we run it on compressed air.  I believe we've
successfully run it on N2 (which fluctuates around 75-80PSI for us) when
our building compressed air is down.

Our building air fluctuates around 103-110PSI (~103 is where the compressor
kicks on), and our N2 (boiloff from a large LN2 storage tank) is around 75-85PSI.   
We use inline particulate filters, oil filters, and regulators before air
goes to any of our magnets.

Best,

Samrat Amin
Magnetic Resonance Research Center
Arizona State University


-=-=-=-=-=-=-=-=-=-=-=-