Re: [AMMRL] AV3 NanoBay IPSO cannot be detected

From: dma6 via groups.io <dma6=uthsc.edu_at_groups.io>
Date: Tue, 27 Oct 2026 21:43:38 -0700

Thank you all who responded to my message! The IPSO can be detected now, but
still needs some troubleshootings. It only boots using the old NMR PC that
came with the AV3 NanoBay system.

I did following to restart the console and change the ethernet switch :
1 Prese the red ON/OFF button on the IPSO for 6 sec to turn off the IPSO.
2 Power off the console from the back cover and let console stay off for at
least 3min
3 While it's off, plug in all 4 red ethernet cables from the original switch
to a new switch.
4 Power the console back on, and wait 10 seconds, (not sure if timing is
important) then turn on the IPSO by briefly pressing the red button

After console restarts, wIth the current Win11 NMR-PC, the IPSO POST CODE
couldn't pass C0. But When I plug the ethernet cable to the old NMR-PC, the
IPSO can pass C0 and manage to boot. The original PC has Windows 7 with TS3.0
The license has expired on it, so I can't start software from the old PC.
But at least I can access the IPSO webpage to get more info.

I still have to figure out how to use the current Windows 11 PC to boot the
IPSO. The firewall on the Win11 PC is always on and set by campus ITS. I think
all the Bruker programs/apps are allowed through firewall. Whenever I check,
Bruker Dhcp server and Tftp server are always on.

Thank you all. It's really helpful. I wish there is one group like AMMRL for
Mass Spec Labs.

Best,
Dejian


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3894): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3894__;!!PvDODwlR4mBZyAb0!S9IJAESjrdUUd0_CSlgJL42q-0kCoT=
CGDbGbOkp8ez6MkGQn5KtqWrFSco_B2t75M4k6x2J8otWe2ouZBxHon5VipQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121921779=
/7559972__;!!PvDODwlR4mBZyAb0!S9IJAESjrdUUd0_CSlgJL42q-0kCoTCGDbGbOkp8ez6Mk=
GQn5KtqWrFSco_B2t75M4k6x2J8otWe2ouZBxGZBxtw3g$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 28 2026 - 02:02:39 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST