Hello Dejian,
I experienced a very similar problem on our AV-III HD nanobay (controlled
by a CentOS 7 workstation) a few months ago.
In our case, the dhcpd service on the host workstation appeared to have
somehow hung or stopped communicating with the IPSO during normal
operation. I rebooted the workstation and the console to resolve the
communication issue, but continued to observe the exact same symptoms you
are seeing: no connection to the IPSO, but everything else was visible in
ha.
After investigating, I found that the dhcpd service wasn't running. Why it
didn't start automatically upon reboot, I do not know. I had to manually
restart the dhcpd service on the workstation using the command line. Once
the dhcpd service was manually restarted, I reset the IPSO, it completed
its normal POST (no hanging on C0), and communication was restored.
Perhaps restart the analogous Windows service(s) and reset the IPSO to see
if it resolves your communication issue?
Best,
Thomas
On Sat, Oct 24, 2026 at 6:18 PM dma6 via groups.io wrote:
> Dear all,
>
> We have an AV3 NanoBay 400. it just happened that the IPSO couldn't be
> detected which might be related to a failed Windows update last week. We
> use TopSpin 3.6.1 on a Windows 11 PC.
>
> 1, ha showed DRU and Lock/Shim/VTU can be accessed But IPSO device "is
> configured in TopSpin but hardware is not accessible.." Both DRU and
> Lock/Shim/VTU webpage can be opened, but the IPSO webpage cannot.
>
> 2, bsmsproxyserver send Heartbeat Check Error from time to time.
>
> 3, cf showed "check hardware there are 0 DHCP controlled devices and 19
> devices with fixed IP to check try to connect 19 devices at the
> spectrometer subnet
> connected: DRU1 at IP 149.236.99.89
> connected: ELCB_... at IP 149.236.99.20
> read configuration data from BSMS/2 "
>
> It hangs here. Another small window showed:
> "cf2 Spectrometer Configuration Error:
> Cannot check hardware: No AV3 spectrometer detected in the spectrometer
> subnet ...
> terminate configuration"
>
> I read previous posts with the same issue, also with help from Bruker
> Support, I tried several things:
> reinstalled the Diskless folder, swapped the CAT5 cables to IPSO and ELCB,
> used a regular hub to replace the switch. all without success. It seems
> that hardware is fine. There is no error light lit in the NanoBay console.
> Power supply and cooling fans are also working as far as I can tell.
>
> I guess it could be Windows DHCP permission issue. Bruker Dhcp Server and
> Tftp Server are both running. Here are my quesitons:
> Is IPSO a DHCP controlled device? In general, which are the DHCP
> controlled devices in NanoBay console?
> Is there a configuration file containing a list of fixed IP devices and
> the DHCP controlled devices?
> Could this be Firewall related although the Bruker apps are allowed
> through firewall?
>
> Thank you for your time.
>
> Dejian
>
> Dejian Ma
> University of Tennessee Health Science Center
> dma6_at_uthsc.edu
>
>
--
Thomas G. Smith
smit3684_at_umn.edu
Chemistry NMR Laboratory <
https://cse.umn.edu/chem/NMR >
University of Minnesota
(612) 625-8374
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3892): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3892__;!!PvDODwlR4mBZyAb0!TPtbX-xPgLPsE1LMbRzNgdQs-VIVku=
fLBAWY4X30xARZrDeVJ9NA5LajUsMZNycVMGTTaY2FksRihVGO8zqVZr4Je3G9$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121921779=
/7559972__;!!PvDODwlR4mBZyAb0!TPtbX-xPgLPsE1LMbRzNgdQs-VIVkufLBAWY4X30xARZr=
DeVJ9NA5LajUsMZNycVMGTTaY2FksRihVGO8zqVZoxyObNW$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Oct 27 2026 - 10:13:56 MST