Re: [AMMRL] AV3 NanoBay IPSO cannot be detected

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Wed, 28 Oct 2026 13:45:06 +0000 (UTC)

 Hi Dejian,

That's a good sign.  So if the IPSO doesn't boot properly from the new
computer but does from the old one then the issue is with something on the
new computer.  If the Bruker DHCP and Bruker TFTP servers are running,
I guess the next thing I would look at is the hosts file.  On a windows
PC that's in C:\Windows\System32\drivers\etc\hosts and if it doesn't have
the two lines:149.236.99.1 ASP_ST2149.236.99.99 spect then it won't correctly
communicate with the console. That's probably the next thing I would check. 
If you end up adding those you should restart the Bruker DHCP and Bruker tftp
servers and then hit the white reset button on the IPSO.  You might also
check both the incoming rules and outgoing rules for the firewall to make sure
that they will allow communication from the console (or even turn off the
firewall for the spectrometer connection if they'll let you.)  

Cheers,Mike

On Wednesday, October 28, 2026 at 03:02:36 AM MDT, dma6 via groups.io wrote:

Thank you all who responded to my message! The IPSO can be detected now, but
still needs some troubleshootings. It only boots using the old NMR PC that
came with the AV3 NanoBay system.

I did following to restart the console and change the ethernet switch :
1 Prese the red ON/OFF button on the IPSO for 6 sec to turn off the IPSO.
2 Power off the console from the back cover and let console stay off for at
least 3min  
3 While it's off, plug in all 4 red ethernet cables from the original switch
to a new switch.
4 Power the console back on, and wait 10 seconds, (not sure if timing is important)
    then turn on the IPSO by briefly pressing the red button
After console restarts, wIth the current Win11 NMR-PC, the IPSO POST CODE
couldn't pass C0. But When I plug the ethernet cable to the old NMR-PC, the
IPSO can pass C0 and manage to boot. The original PC has Windows 7 with TS3.0
The license has expired on it, so I can't start software from the old PC.
But at least I can access the IPSO webpage to get more info.

I still have to figure out how to use the current Windows 11 PC to boot the
IPSO. The firewall on the Win11 PC is always on and set by campus ITS. I think
all the Bruker programs/apps are allowed through firewall. Whenever I check,
Bruker Dhcp server and Tftp server are always on.

Thank you all. It's really helpful. I wish there is one group like AMMRL for
Mass Spec Labs.

Best,Dejian


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3895): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3895__;!!PvDODwlR4mBZyAb0!TakvAbHm3Cuy9nnorRsLImQVee2QG9=
YdPA7RmPfMXsjC1lHwTrAQ2mhdUt871deQKC0BwTVGaREEm61hawN0CQiGATTH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121921779=
/7559972__;!!PvDODwlR4mBZyAb0!TakvAbHm3Cuy9nnorRsLImQVee2QG9YdPA7RmPfMXsjC1=
lHwTrAQ2mhdUt871deQKC0BwTVGaREEm61hawN0CQ0xw4wU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 28 2026 - 06:59:32 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST