From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Mon, 26 Oct 2026 09:52:24 +0100
Hi,

did you already try to 
-disconnect any external ethernet connection from the PC (I saw it once that due to an incomplete
firewall the PC found an external server under the address used by the IPSO)
- switch off the firewall completely (external ethernet unplugged) just to find out whether the
firewall doesn’t let pass IPSO data packets

Please post any solution you find…

Best regards 
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



Am 25.10.2026 um 01:18 schrieb dma6 via groups.io <dma6=uthsc.edu_at_groups.io>:

Dear all,

We have an AV3 NanoBay 400. it just happened that the IPSO couldn't be detected which might be related to a failed Windows update last week.  We use TopSpin 3.6.1 on a Windows 11 PC.

1, ha showed DRU and Lock/Shim/VTU can be accessed But IPSO device "is configured in TopSpin but hardware is not accessible.."  Both DRU and Lock/Shim/VTU webpage can be opened, but the IPSO webpage cannot.

2, bsmsproxyserver send Heartbeat Check Error from time to time.

3, cf showed "check hardware there are 0 DHCP controlled devices and 19 devices with fixed IP to check try to connect 19 devices at the spectrometer subnet
connected: DRU1 at IP 149.236.99.89
connected: ELCB_... at IP 149.236.99.20
read configuration data from BSMS/2 "

It hangs here. Another small window showed:
"cf2 Spectrometer Configuration Error:
Cannot check hardware: No AV3 spectrometer detected in the spectrometer subnet ...
terminate configuration"

I read previous posts with the same issue, also with help from Bruker Support, I tried several things:
reinstalled the Diskless folder, swapped the CAT5 cables to IPSO and ELCB, used a regular hub to replace the switch.  all without success.  It seems that hardware is fine. There is no error light lit in the NanoBay console.  Power supply and cooling fans are also working as far as I can tell.
 
I guess it could be Windows DHCP permission issue.  Bruker Dhcp Server and Tftp Server are both running.  Here are my quesitons:
Is IPSO a DHCP controlled device? In general, which are the DHCP controlled devices in NanoBay console?
Is there a configuration file containing a list of fixed IP devices and the DHCP controlled devices? 
Could this be Firewall related although the Bruker apps are allowed through firewall? 
 
Thank you for your time. 
 
Dejian
 
Dejian Ma
University of Tennessee Health Science Center
dma6_at_uthsc.edu
 
 
_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#3885) | Reply = to Group | Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Mon Oct 26 2026 - 02:45:36 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST