Re: [AMMRL] AV3 NanoBay IPSO cannot be detected

From: Ping Yu via groups.io <pyu=ucdavis.edu_at_groups.io>
Date: Mon, 26 Oct 2026 04:16:50 +0000

Hi Dejian,

We had similar issues a couple of months ago. It turned out that the ethernet
switch board inside the console was bad. The aged ethernet switch board caused
communication issues between the console and the probe. We put a new external
ethernet switch box to replace the old one. All the cables on the old ethernet
board were connected to the new one. We left the old one there. The console space
is just enough to accommodate the new ethernet board.

We were able to find path to let the power cord reach a power outlet near the console.

Good luck.

Ping

> From: main_at_ammrl.groups.io on behalf of dma6 via groups.io
> Date: Saturday, October 24, 2026 at 4:18 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] AV3 NanoBay IPSO cannot be detected

Dear all,

We have an AV3 NanoBay 400. it just happened that the IPSO couldn't be detected
which might be related to a failed Windows update last week. We use TopSpin 3.6.1
on a Windows 11 PC.

1, ha showed DRU and Lock/Shim/VTU can be accessed But IPSO device "is configured
in TopSpin but hardware is not accessible.." Both DRU and Lock/Shim/VTU webpage
can be opened, but the IPSO webpage cannot.

2, bsmsproxyserver send Heartbeat Check Error from time to time.

3, cf showed "check hardware there are 0 DHCP controlled devices and 19 devices
with fixed IP to check try to connect 19 devices at the spectrometer subnet
connected: DRU1 at IP 149.236.99.89
connected: ELCB_... at IP 149.236.99.20
read configuration data from BSMS/2 "

It hangs here. Another small window showed:
"cf2 Spectrometer Configuration Error:
Cannot check hardware: No AV3 spectrometer detected in the spectrometer subnet ...
terminate configuration"

I read previous posts with the same issue, also with help from Bruker Support,
I tried several things:
reinstalled the Diskless folder, swapped the CAT5 cables to IPSO and ELCB,
used a regular hub to replace the switch. all without success. It seems
that hardware is fine. There is no error light lit in the NanoBay console.
Power supply and cooling fans are also working as far as I can tell.

I guess it could be Windows DHCP permission issue. Bruker Dhcp Server and
Tftp Server are both running. Here are my quesitons:
Is IPSO a DHCP controlled device? In general, which are the DHCP controlled
devices in NanoBay console?
Is there a configuration file containing a list of fixed IP devices and the
DHCP controlled devices?
Could this be Firewall related although the Bruker apps are allowed through
firewall?

Thank you for your time.

Dejian

Dejian Ma
University of Tennessee Health Science Center
dma6_at_uthsc.edu





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3884): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3884__;!!PvDODwlR4mBZyAb0!UWQROj7ygUER_ekOtm145UkNwHkNCu=
YpbbnhUI-GRkmhcmshhYA0lyLqdlr20Zmz_S9fZ-VSJBzPE0K096rKbd2iKas$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121921779=
/7559972__;!!PvDODwlR4mBZyAb0!UWQROj7ygUER_ekOtm145UkNwHkNCuYpbbnhUI-GRkmhc=
mshhYA0lyLqdlr20Zmz_S9fZ-VSJBzPE0K096rK8R015fA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 26 2026 - 02:45:17 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST