Re: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

From: Paris Ning, PhD via groups.io <pning=isotope.com_at_groups.io>
Date: Thu, 15 Oct 2026 19:39:09 +0000

Hi all,

As someone having worked exclusively on CMP-NMR for my PhD, I’d suggest
using the screw plug insert for HRMAS!

This is Bruker’s link (https://store.bruker.com/products/hrmas-insert-kitbut1 )
but I figured you may be able to find something cheaper else where. I was able to reuse
each multiple times. However, with solvent inside the sample matrix, the spin rate
becomes an issue. I was able to get to 9000 with a quite slushy/wet sample but I
have not dealt with “physically robust gel” samples so your milage may vary.

Hope this helps!

Cheers,
Paris

> From: main_at_ammrl.groups.io On Behalf Of Ulrike Werner-Zwanziger via groups.io
> Sent: Thursday, October 15, 2026 3:07 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

Hello,

I second rotor inserts. We purchased some from cortec many years ago. They
are a plastic container with a plug pushed in. That plug has a thread in the
top, to pull the whole assembly out of the (4mm) rotor. Unfortunately, the
containers cannot be reused, but they are leak proof and no mess afterwards.
They spin very stably.

In the past we also used glass inserts which contain a narrow glass part and
a funnel on top. You have to seal or superglue and cut the narrow part. Again
a one time deal. They were difficult to remove from the rotor.

Happy resonating (and spinnig)

Ulli


Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way
PO BOX 15000
Halifax, NS B3H 4R2

e-mail: Ulli.Zwanziger_at_Dal.ca<mailto:Ulli.Zwanziger_at_Dal.ca>

Web: http://www.dal.ca/diff/nmr3.html
Tel. 1 902 494 8085

FAX: 1 902 494 1310



On Thu, Oct 15, 2026, 17:39 Nieuwendaal, Ryan C. (Fed) via groups.io wrote:
Hi Ken,
I would recommend using the rotor inserts. I use them and really like them
for the 4mm. That makes them easy to switch out, the seals are good, and the
user can even just purchase several for their samples and you can just use
the same rotor over and over.

Ryan

> From: main_at_ammrl.groups.io On Behalf Of Michael Groves via groups.io
> Sent: Thursday, October 15, 2026 10:19 AM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [EXTERNAL] Re: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

Hi Ken,

My one recommendation is:

If you're concerned about getting the stuff out of the rotor afterwards, just
have them buy their own rotor(s) and then they can deal with it.

Cheers,
Mike

On Thursday, October 15, 2026 at 07:45:47 AM MDT, Kenneth Sharp-Knott via groups.io wrote:

Hi everyone,
We have a user asking to run gel samples on our 400 MHz system equipped with
a 4mm SSNMR probe. These are battery-related materials, and the student describes
the sample as a "physically robust" gel.
Even with a firmer gel, the idea of spinning trapped liquid or volatile solvents
in our solids stator makes me nervous. Before we let them run, I’d
love to hear what practical experience others have had with this kind of material:

  * What cap or sealing setup do you recommend to make sure the solvent doesn't
leak out under centrifugal load?
  * What upper limits do you enforce on MAS spinning speeds to keep things
stable and avoid forcing liquid out of the gel? We were thinking of keeping
it around 4–6 kHz.
  * What loading tricks or methods work best to get sticky or rubbery gels
packed evenly without trapping air?
 * Any tips on getting the gel out afterward and cleaning the rotor safely?
Any advice, warnings, or "lessons learned the hard way" would be greatly appreciated.
Thanks,
Ken Sharp-Knott
Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3871): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3871__;!!PvDODwlR4mBZyAb0!Vr5cALDBv0a_dX_UHyIlC7oSJYgcUi=
N4kG9JrKc1PJ45u4shWECnJynJ1lpVY_YWDxDhQ0AeL98RzkiOOpcZJOzPKdAN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!Vr5cALDBv0a_dX_UHyIlC7oSJYgcUiN4kG9JrKc1PJ45u=
4shWECnJynJ1lpVY_YWDxDhQ0AeL98RzkiOOpcZJPlBbnUJ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






image001.png
(image/png attachment: image001.png)

Received on Thu Oct 15 2026 - 12:51:08 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST