Hello,
I second rotor inserts. We purchased some from cortec many years ago. They
are a plastic container with a plug pushed in. That plug has a thread in
the top, to pull the whole assembly out of the (4mm) rotor. Unfortunately,
the containers cannot be reused, but they are leak proof and no mess
afterwards. They spin very stably.
In the past we also used glass inserts which contain a narrow glass part
and a funnel on top. You have to seal or superglue and cut the narrow part.
Again a one time deal. They were difficult to remove from the rotor.
Happy resonating (and spinnig)
Ulli
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way
PO BOX 15000
Halifax, NS B3H 4R2
e-mail: Ulli.Zwanziger_at_Dal.ca
Web: *http://www.dal.ca/diff/nmr3.html*
Tel. 1 902 494 8085
FAX: 1 902 494 1310
On Thu, Oct 15, 2026, 17:39 Nieuwendaal, Ryan C. (Fed) via groups.io wrote:
> Hi Ken,
>
> I would recommend using the rotor inserts. I use them and really like them
> for the 4mm. That makes them easy to switch out, the seals are good, and
> the user can even just purchase several for their samples and you can just
> use the same rotor over and over.
>
> Ryan
>
>
>
> *From:* main_at_ammrl.groups.io <main_at_ammrl.groups.io> *On Behalf Of *Michael> Groves via groups.io
> *Sent:* Thursday, October 15, 2026 10:19 AM
> *To:* main_at_ammrl.groups.io
> *Subject:* [EXTERNAL] Re: [AMMRL] Seeking practical advice for running> battery gel electrolytes on a 4mm SSNMR probe?
>
> Hi Ken,
>
> My one recommendation is:
>
> If you're concerned about getting the stuff out of the rotor afterwards,
> just have them buy their own rotor(s) and then they can deal with it.
>
> Cheers,
>
> Mike
>
> On Thursday, October 15, 2026 at 07:45:47 AM MDT, Kenneth Sharp-Knott via> groups.io wrote:
>
>
>
>
>
> Hi everyone,
>
> We have a user asking to run gel samples on our 400 MHz system equipped
> with a 4mm SSNMR probe. These are battery-related materials, and the
> student describes the sample as a "physically robust" gel.
>
> Even with a firmer gel, the idea of spinning trapped liquid or volatile
> solvents in our solids stator makes me nervous. Before we let them run, I’d
> love to hear what practical experience others have had with this kind of
> material:
>
> - What cap or sealing setup do you recommend to make sure the solvent
> doesn't leak out under centrifugal load?
> - What upper limits do you enforce on MAS spinning speeds to keep
> things stable and avoid forcing liquid out of the gel? We were thinking of
> keeping it around 4–6 kHz.
> - What loading tricks or methods work best to get sticky or rubbery
> gels packed evenly without trapping air?
> - Any tips on getting the gel out afterward and cleaning the rotor
> safely?
>
> Any advice, warnings, or "lessons learned the hard way" would be greatly
> appreciated.
>
> Thanks,
>
> Ken Sharp-Knott
>
> Senior Manager, NMR & Analytical Core Facilities
> Department of Chemistry
> Virginia Tech
>
> (540)267-6502 (Cell)
> (540)231-0885 (Office)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3870): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3870__;!!PvDODwlR4mBZyAb0!Tn-bg6ScspkX9LBCsy7lGlwPtbjY4z=
m6aKk394RuPzTcx6JB-Uw5uX5RdmjpokRAvXRXlRWsRuR5KiiSQ8YfvkJN8ZZQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!Tn-bg6ScspkX9LBCsy7lGlwPtbjY4zm6aKk394RuPzTcx=
6JB-Uw5uX5RdmjpokRAvXRXlRWsRuR5KiiSQ8Yfvqpu9C5F$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 15 2026 - 12:07:31 MST