Hi Ken,
I would recommend using the rotor inserts. I use them and really like them
for the 4mm. That makes them easy to switch out, the seals are good, and the
user can even just purchase several for their samples and you can just use
the same rotor over and over.
Ryan
> From: main_at_ammrl.groups.io On Behalf Of Michael Groves via groups.io
Sent: Thursday, October 15, 2026 10:19 AM
To: main_at_ammrl.groups.io
Subject: [EXTERNAL] Re: [AMMRL] Seeking practical advice for running batter=
y gel electrolytes on a 4mm SSNMR probe?
Hi Ken,
My one recommendation is:
If you're concerned about getting the stuff out of the rotor afterwards, just
have them buy their own rotor(s) and then they can deal with it.
Cheers,
Mike
On Thursday, October 15, 2026 at 07:45:47 AM MDT, Kenneth Sharp-Knott via groups.io wrote:
Hi everyone,
We have a user asking to run gel samples on our 400 MHz system equipped with a
4mm SSNMR probe. These are battery-related materials, and the student describes
the sample as a "physically robust" gel.
Even with a firmer gel, the idea of spinning trapped liquid or volatile solvents
in our solids stator makes me nervous. Before we let them run, I'd love
to hear what practical experience others have had with this kind of material:
* What cap or sealing setup do you recommend to make sure the solvent doesn't
leak out under centrifugal load?
* What upper limits do you enforce on MAS spinning speeds to keep things
stable and avoid forcing liquid out of the gel? We were thinking of keeping
it around 4-6 kHz.
* What loading tricks or methods work best to get sticky or rubbery gels
packed evenly without trapping air?
* Any tips on getting the gel out afterward and cleaning the rotor safely?
Any advice, warnings, or "lessons learned the hard way" would be greatly appreciated.
Thanks,
Ken Sharp-Knott
Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech
(540)267-6502 (Cell)
(540)231-0885 (Office)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3868): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3868__;!!PvDODwlR4mBZyAb0!XYsOAnn8XMtJQ8V0iLws7-gAo5swQP=
KLS8ayDwX7R78qEreJZNJc-TzraqJNPmieS5TaYQRkVjqtXJ6q9GArYRmoaiEa$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!XYsOAnn8XMtJQ8V0iLws7-gAo5swQPKLS8ayDwX7R78qE=
reJZNJc-TzraqJNPmieS5TaYQRkVjqtXJ6q9GArYegEzP01$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 15 2026 - 08:39:08 MST