Hi Ken,
my one recommendation is:
If you're concerned about getting the stuff out of the rotor afterwards, just
have them buy their own rotor(s) and then they can deal with it.
Cheers,Mike
On Thursday, October 15, 2026 at 07:45:47 AM MDT, Kenneth Sharp-Knott wrote:
Hi everyone,We have a user asking to run gel samples on our 400 MHz system
equipped with a 4mm SSNMR probe. These are battery-related materials, and
the student describes the sample as a "physically robust" gel.Even with a
firmer gel, the idea of spinning trapped liquid or volatile solvents in our
solids stator makes me nervous. Before we let them run, I’d love to
hear what practical experience others have had with this kind of material:
- What cap or sealing setup do you recommend to make sure the solvent doesn't
leak out under centrifugal load?
- What upper limits do you enforce on MAS spinning speeds to keep things
stable and avoid forcing liquid out of the gel? We were thinking of keeping
it around 4–6 kHz.
- What loading tricks or methods work best to get sticky or rubbery gels
packed evenly without trapping air?
- Any tips on getting the gel out afterward and cleaning the rotor safely?
Any advice, warnings, or "lessons learned the hard way" would be greatly
appreciated.Thanks,Ken Sharp-KnottSenior Manager, NMR & Analytical Core Facilities
Department of Chemistry Virginia Tech
(540)267-6502 (Cell)(540)231-0885 (Office)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3867): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3867__;!!PvDODwlR4mBZyAb0!SbAwqFGiOs_7w99oKGrvwthdTWAAQv=
caEtwn-eX7-TeFQ6OD26XJDLVC-m-SjXYoaegvrDIwDiqdLbJjxIJwt1J-VKIj$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!SbAwqFGiOs_7w99oKGrvwthdTWAAQvcaEtwn-eX7-TeFQ=
6OD26XJDLVC-m-SjXYoaegvrDIwDiqdLbJjxIJwt7J6p-91$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 15 2026 - 08:01:59 MST