Dear Ken,
this kind of semi-solid samples, like gels, tissues, swellable
resins,.. are the typical samples for HR-MAS NMR.
There are special HR-MAS rotors available which have a inner volume of
either 12ul or 50 ul.
If you press in the upper part of the spacer, which has a small threaded
hole, air and excess solvent can get out. You than remove the solvent or
gel with a kleenex or small spatula.
The hole is sealed with a small screw.
https://store.bruker.com/products/hrmas-rotor-insert-kit
Test your spinning speed stepwise starting at 1 kHz up to 6 or even 10
kHz and check for sample stability and line narrowing. It strongly
depends of the typ of sample.
Sample preparation can be a mess for sticky samples. To get samples out
I used a small spatula, afterwards a ultrasonic bath.
Best wishes from Heidelberg
Jürgen
--
Dr. Jürgen Graf
Organisch-Chemisches Institut
Kernresonanzspektroskopie
Ruprecht-Karls-Universität Heidelberg
Im Neuenheimer Feld 270
69120 Heidelberg
Telefon: 06221 / 54-8431
E-Mail:graf_at_oci.uni-heidelberg.de
Am 15.10.2026 um 15:28 schrieb Kenneth Sharp-Knott via groups.io:
> Hi everyone,
> We have a user asking to run gel samples on our 400 MHz system
> equipped with a 4mm SSNMR probe. These are battery-related materials,
> and the student describes the sample as a "physically robust" gel.
> Even with a firmer gel, the idea of spinning trapped liquid or
> volatile solvents in our solids stator makes me nervous. Before we let
> them run, I’d love to hear what practical experience others have had
> with this kind of material:
>
> *
> What cap or sealing setup do you recommend to make sure the
> solvent doesn't leak out under centrifugal load?
> *
> What upper limits do you enforce on MAS spinning speeds to keep
> things stable and avoid forcing liquid out of the gel? We were
> thinking of keeping it around 4–6 kHz.
> *
> What loading tricks or methods work best to get sticky or rubbery
> gels packed evenly without trapping air?
> *
> Any tips on getting the gel out afterward and cleaning the rotor
> safely?
>
> Any advice, warnings, or "lessons learned the hard way" would be
> greatly appreciated.
> Thanks,
> Ken Sharp-Knott
> Senior Manager, NMR & Analytical Core Facilities
> Department of Chemistry
> Virginia Tech
>
> (540)267-6502 (Cell)
> (540)231-0885 (Office)
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3865): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3865__;!!PvDODwlR4mBZyAb0!Xwp2iU03S8kv5Z-LFXaLP7xlXy3sEB=
9eAtmFaVM7DpdzErLUL6m9s-c3Kv_Va0n5MB7HMhJTANgc7QfhyGiPHUks1I7J$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!Xwp2iU03S8kv5Z-LFXaLP7xlXy3sEB9eAtmFaVM7DpdzE=
rLUL6m9s-c3Kv_Va0n5MB7HMhJTANgc7QfhyGiPHZSIvzY5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 15 2026 - 07:59:26 MST