Re: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

From: Alex Paterson via groups.io <alpaterson=wisc.edu_at_groups.io>
Date: Thu, 15 Oct 2026 14:10:03 +0000

Hi Ken,

I’m also working with a battery group on a “physically robust” gel.
In my case it’s more gel-like after heating, but similar principles apply.

For sealing, you can purchase 4 mm caps that have O rings. In my experience
Bruker 4 mm rotors with O-ring caps seal exceptionally well, good enough for
in situ measurements of air-sensitive chemistries. Should handle solvents
without issue. In the absence of those, you can follow guides for spinning
biological samples, which typically involves pressing a rubber disc between
the sample and the cap. You also have the potential to use a HRMAS insert
kit, which may be particularly useful for a soupy sample.

  1.
  2. Standard rotor rating minus about 10%. Big thing is to start slow and
check for stability at lower speeds before going to high speeds. If you have
a spin test station it would be helpful to spin there first.
  3. Small portions and a lot of patience. You probably will trap some air;
this is why starting at lower speeds is important, give things a chance to
redistribute.
  4. Unfortunately this one I can’t help with. My SOP for cleaning
difficult samples is only good for Phoenix rotors, where you can easily remove
both the top and bottom caps and have an empty zirconia sleeve. I’d
ask the user if they have a compatible solvent that would work with
loosening the gel.

Now, all of the above is with a sample that packs like a powder, but melts
in the rotor above about 115 degC; so I melt while spinning, then unpack the
material afterwards. If your sample is more soupy you’ll want to
start slower.

However, in general, slow and steady should be fine.

Good luck!
Alex

Alex Paterson
The National Magnetic Resonance Facility at Madison (NMRFAM)
University of Wisconsin-Madison
alpaterson_at_wisc.edu


> From: main_at_ammrl.groups.io On Behalf Of Kenneth Sharp-Knott via groups.io
> Sent: Thursday, October 15, 2026 8:28 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

Hi everyone,

We have a user asking to run gel samples on our 400 MHz system equipped with
a 4mm SSNMR probe. These are battery-related materials, and the student describes
the sample as a "physically robust" gel.

Even with a firmer gel, the idea of spinning trapped liquid or volatile solvents
in our solids stator makes me nervous. Before we let them run, I’d love
to hear what practical experience others have had with this kind of material:

  * What cap or sealing setup do you recommend to make sure the solvent doesn't
leak out under centrifugal load?
  * What upper limits do you enforce on MAS spinning speeds to keep things
stable and avoid forcing liquid out of the gel? We were thinking of keeping
it around 4–6 kHz.
  * What loading tricks or methods work best to get sticky or rubbery gels
packed evenly without trapping air?
  * Any tips on getting the gel out afterward and cleaning the rotor safely?
Any advice, warnings, or "lessons learned the hard way" would be greatly appreciated.

Thanks,

Ken Sharp-Knott

Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3864): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3864__;!!PvDODwlR4mBZyAb0!Q86GL_lM3anP4LiUfUZFnVoBdh_vMW=
6H2VuuzfpYYGbToM_mXfLl9VYcp4PB5LgCugMWZM3iicqIV24zspUwnot5RY8h$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!Q86GL_lM3anP4LiUfUZFnVoBdh_vMW6H2VuuzfpYYGbTo=
M_mXfLl9VYcp4PB5LgCugMWZM3iicqIV24zspUwnggIqfhT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Oct 15 2026 - 07:59:02 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST