Re: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

From: D'Anciaes Almeida Silva, Igor via groups.io <idhbx=missouri.edu_at_groups.io>
Date: Thu, 15 Oct 2026 14:06:03 +0000

Hi Ken,

To load the gel into the rotor you could try a centrifuge (I used to do it
for microgels). To get the sample out, we used to use a solvent that could
dissolve the sample, but even with this is not easy. The viscous the gel
the harder is.

For rotation, I used 10 kHz for these microgels and had no issues with
leaking using the regular setup.

Hope I helped.

----------------
Igor d'Ancićes Almeida Silva, Ph.D.
Director - Chemistry NMR Facility
University of Missouri-Columbia
32A Chemistry Building
________________________________
> From: main_at_ammrl.groups.io on behalf of Kenneth Sharp-Knott via groups.io
> Sent: Thursday, October 15, 2026 8:28 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Seeking practical advice for running battery gel electrolytes on a 4mm SSNMR probe?

Hi everyone,

We have a user asking to run gel samples on our 400 MHz system equipped with
a 4mm SSNMR probe. These are battery-related materials, and the student
describes the sample as a "physically robust" gel.

Even with a firmer gel, the idea of spinning trapped liquid or volatile solvents
in our solids stator makes me nervous. Before we let them run, I’d love
to hear what practical experience others have had with this kind of
material:

  *
What cap or sealing setup do you recommend to make sure the solvent doesn't
leak out under centrifugal load?
  *
What upper limits do you enforce on MAS spinning speeds to keep things stable
and avoid forcing liquid out of the gel? We were thinking of keeping it
around 4–6 kHz.
  *
What loading tricks or methods work best to get sticky or rubbery gels packed
evenly without trapping air?
  *
Any tips on getting the gel out afterward and cleaning the rotor safely?

Any advice, warnings, or "lessons learned the hard way" would be greatly
appreciated.

Thanks,

Ken Sharp-Knott

Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3863): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3863__;!!PvDODwlR4mBZyAb0!WShk7jBlL7zJY592v7qqbhGoajQZRQ=
mF5kodCDuU0RrnxpyVB0hhPy8QVsxPxQVyST9S5LHcBjgW8Q9DKWdjfw2LZ5el$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121769966=
/7559972__;!!PvDODwlR4mBZyAb0!WShk7jBlL7zJY592v7qqbhGoajQZRQmF5kodCDuU0Rrnx=
pyVB0hhPy8QVsxPxQVyST9S5LHcBjgW8Q9DKWdjfxX5WPhe$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Oct 15 2026 - 07:58:13 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST