[AMMRL] How to determine safe spin lock power/duration

From: Brian Michael via groups.io <bmichael=brandeis.edu_at_groups.io>
Date: Thu, 15 Oct 2026 15:46:36 -0400

Hello,

A user wants to do some 19F T1rho measurements and I'm trying to figure out
safe limits to tell them for the power/duration of the spin lock pulse. The
specs I have from Bruker for 19F on our cryoprobe give limits of 1ms at the
power level of a 11us 90 pulse for a hard pulse and 300ms at the power
level of a 30us 90 pulse for GARP decoupling.

I'm wondering how others usually go about determining safe durations for
power levels in between those listed in the specs. How long would it be
safe to apply a pulse at the power level of a 16us or 20us 90 pulse?

Thanks,
Brian Michael

NMR Research Facility Director
Brandeis University


-=-=-=-=-=-=-=-=-=-=-=-