Hi All,
I would first ask whether you experienced a power outage? If so, or in any
event, open the lock screen gui and check to be sure the lock phase is adjusted
correctly. ..if the phase is off, it may not reliably find the Z0. Can you
lock it manually in the gui, or when you get the lock on resonance, and check
the lock box, it shifts out, you most likely have issues to be determined,
given no other circumstances.