Hi,

Try to move the z0 slider during Lock Scan and observe whether you can change
the offset frequency. If not, check:
- is the shim cable still properly connected on both ends
- is the shim power supply switched on, any blown fuses on the back
- which shim power supply do you have? Large blank aluminium front or painted
beige front? Might be the z0z1 board

Cheers