Hi John:
As far as I knew, we didn’t have a power outage over the weekend.
I could lock manually but the lock signal was not stable and kept changing.
Thanks,
Nelson
> From: main_at_ammrl.groups.io On Behalf Of johnhahn via groups.io
> Sent: Monday, October 6, 2025 5:46 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] find z0 problem on INOVA 400
Hi All,
I would first ask whether you experienced a power outage? If so, or in any
event, open the lock screen gui and check to be sure the lock phase is adjusted
correctly. ..if the phase is off, it may not reliably find the Z0. Can you lock
it manually in the gui, or when you get the lock on resonance, and check the
lock box, it shifts out, you most likely have issues to be determined, given
no other circumstances.
Regards,
John E. Hahn
CAZ Enterprises, LLC
6320 Fly Rd. Ste 207
East Syracuse, NY 13057
Ph. 315-727-8192
http://www.Cazent.com
-=-=-=-=-=-=-=-=-=-=-=-