Re: [AMMRL] Tuning rods pinned?

From: Karel Klika via groups.io <karkli=utu.fi_at_groups.io>
Date: Thu, 2 Oct 2025 13:15:39 +0000

Hi Jim,

I am not sure if I get exactly the same error message, but I get something
similar nowadays whenever I reboot TopSpin and try to do atma or atmm for the
first time. The only way I can seem to clear the error is to repeatedly
cycle through wobb and then atmm. It usually only takes two iterations till
I see the dip in wobb and then I know atma/atmm will then work and continue
to work (at least till the next reboot).

Thanks Clemens, I will try that next time!

Regards,
K. Klika
________________________________________
> From: Baleja, James D via groups.io
> Sent: 02 October 2025 00:53:29
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Tuning rods pinned?

Hi all, I'm getting this message, which I think means that the tuning rods=
are pinned. I can tune fine on 13C.
atma_ProbeCorba_impl::start exception: Error: channel can't be wobbled with
this probe [Probe: [1H] can't be selected on the [1H] channel [Nuc-MotorBa=
se:movePreset-Motor-Nr:3 <1H Tune>: can't move to position -244 (current:3)]]

This is on a Bruker 600 Avance Probe TXI 600S3 H-C/N-D-05 Z

I know that a user was trying to tune the 1H channel to 19F, which I don't
think this probe will do.

Any comments would be welcome!

Thanks

Jim




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2869): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2869__;!!PvDODwlR4mBZyAb0!S5mfPVbUV4VBSRucGtOvizdjyv5NDf=
sQctQFGXCnatNGU-NoJYLeVzMBokL3OE5BfYW5CHY70BFGuFQCeYmsotBI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115545023=
/7559972__;!!PvDODwlR4mBZyAb0!S5mfPVbUV4VBSRucGtOvizdjyv5NDfsQctQFGXCnatNGU=
-NoJYLeVzMBokL3OE5BfYW5CHY70BFGuFQCecR9MAIu$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Thu Oct 02 2025 - 06:16:30 MST

This archive was generated by hypermail 2.4.0 : Sat Nov 01 2025 - 15:23:01 MST