Hello Jim,

I have seen this error message with a few probes. Each case it was a failure
(or start of a failure) to a ATM motor in the probe.

Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry



On Oct 1, 2025, at 5:53 PM, Baleja, James D via groups.io wrote:

Hi all, I'm getting this message, which I think means that the tuning rods
are pinned. I can tune fine on 13C.
atma_ProbeCorba_impl::start exception: Error: channel can't be wobbled with
this probe [Probe: [1H] can't be selected on the [1H] channel [Nuc-MotorBa=
se:movePreset-Motor-Nr:3 <1H Tune>: can't move to position -244 (current:3)]]

This is on a Bruker 600 Avance Probe TXI 600S3 H-C/N-D-05 Z

I know that a user was trying to tune the 1H channel to 19F, which I don't
think this probe will do.

Any comments would be welcome!

Thanks

Jim

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2873): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2873__;!!PvDODwlR4mBZyAb0!Vdl4evI9KUurISWhbHaQSVNDF2PZsn=
rXlIReGa4f17J6ElQCxVuvn-b6YWTSNJ6HRyYc94CgQ1Q7atkS4JwQL3shB9vL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115545023=
/7559972__;!!PvDODwlR4mBZyAb0!Vdl4evI9KUurISWhbHaQSVNDF2PZsnrXlIReGa4f17J6E=
lQCxVuvn-b6YWTSNJ6HRyYc94CgQ1Q7atkS4JwQL0bhitZq$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Oct 03 2025 - 03:48:35 MST

This archive was generated by hypermail 2.4.0 : Sat Nov 01 2025 - 15:23:01 MST