you have to consider the type of probes when switching gases. Probes can be
compensated for air for for N2. The oxygen in air changes the susceptibility
quite a bit. Most probes at 400 MHz are air compensated and all probes >=
500MHz are N2 compensated.
Lineshape and resolution specs are no longer guaranteed when using different
gas. You can still get good shims but performance for water suppression might
suffer.
It was already mentioned that if you had been using a cartridge that was
removing oxygen from you house air. Usually a grey cylinder inside or behind
the console then you can remove it after switching to nitrogen.
Regards
Clemens
On 9/25/2025 6:22 PM, Fu Chen via groups.io wrote:
Hello all,
We are considering replacing the compressed air with compressed house N2.
Does anyone have a spec sheet or other documents for comparing using
compressed N2 gas vs compressed Air for Bruker NMR instruments?
Thanks,
Fu
Fu Chen, PhD
Director of Analytical NMR Facility
Department of Chemistry and Biochemistry
B0208A Chemistry Building
University of Maryland, College Park, MD 20742
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2868): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2868__;!!PvDODwlR4mBZyAb0!Rw1UIWCEopQThQ7E_016-OWs8VrZIm=
MtUUY-0ZvhbwsLJ3akAliwhcmG8VzPOo8dQmphxLd_c7Nqdce1KxnlHVDXd6bM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115447661=
/7559972__;!!PvDODwlR4mBZyAb0!Rw1UIWCEopQThQ7E_016-OWs8VrZImMtUUY-0ZvhbwsLJ=
3akAliwhcmG8VzPOo8dQmphxLd_c7Nqdce1KxnlHTdwx2AL$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Oct 02 2025 - 06:16:23 MST