HI Emily
If things are still being installed no worries. Before you hook the magnets
up you are going to have to seal the system and check it for leaks. Depending
on how you do this you are going to do several pumps and purges before you
hook the magnets up. If all the equipment is not there but the plumbing is
done, and the leak check is done, then just let the system purge through the
plumbing until the equipment arrives. We had our plumbing installed months
before the rest of the system arrived. So we let boil-off and magnet fill
gas purge the system. This also gave use a chance to test system pressures to
ensure no system had back pressure during fills.
Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX 77843-3255
gpwylie_at_...
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1734): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1734__;!!PvDODwlR4mBZyAb0!XohxOWkyMXWCgqDzdaC4xy0E6_uP-h=
LjNymPy9aLsdNM3SEINMWJt7ZyIMMEvatiI6XqbOa-vl0ixcyTy8p9zGy_AEVJ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108875353=
/7559972__;!!PvDODwlR4mBZyAb0!XohxOWkyMXWCgqDzdaC4xy0E6_uP-hLjNymPy9aLsdNM3=
SEINMWJt7ZyIMMEvatiI6XqbOa-vl0ixcyTy8p9zOTJKnZz$
Mute #heliumrecovery:
https://urldefense.com/v3/__https://ammrl.groups.io/g/=
main/mutehashtag/heliumrecovery__;!!PvDODwlR4mBZyAb0!XohxOWkyMXWCgqDzdaC4xy=
0E6_uP-hLjNymPy9aLsdNM3SEINMWJt7ZyIMMEvatiI6XqbOa-vl0ixcyTy8p9zJfuiooJ$
Mute #question:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/question__;!!PvDODwlR4mBZyAb0!XohxOWkyMXWCgqDzdaC4xy0E6_uP-hLjNy=
mPy9aLsdNM3SEINMWJt7ZyIMMEvatiI6XqbOa-vl0ixcyTy8p9zEDBmoAS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Oct 07 2024 - 13:11:25 MST