Weigou,
Please check the settings in the probe configuration and see if MAS eject is
turned on. Check also edhead and see if it is set to eject air.
Good luck
> From: main_at_ammrl.groups.io On Behalf Of weiguo.hu via groups.io
> Sent: Monday, October 7, 2024 1:50 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AMMRL: Bruker MAS probe rotor ejection problem
Mike: I disconnected the eject air from the probe and tried eject four times
and every time the eject air turned on. So I guess the problem is indeed inside
the probe.
Brian: I put on an o-ring around the neck of the transport tube (I found that
was the only place where an o-ring can stay in a stable position) so that the
tube is sitting 1-2 mm higher, and unfortunately the eject air didn't turn on.
Any further suggestions would be greatly appreciated. I could send the probe
to Bruker for a TLC but we have a project now that really needs the probe so it
would be great if it can be addressed in-house.
Thanks,
Weiguo
-=-=-=-=-=-=-=-=-=-=-=-