Re: [AMMRL] Limit access to Temp. Display Bruker #Bruker #NMR_Aquisition_Software #NMR_Liquids

From: Struppe, Jochem via groups.io <Jochem.Struppe=bruker.com_at_groups.io>
Date: Thu, 3 Oct 2024 13:46:25 +0000

Greg,

Temperature limits are typically transferred to the temperature control system
by the probe table – in edhead or edprobe. These limits are typically
part of the PICS chip on probes. What you can and should do in theconfiguration
menu Thusitha showed , is to limit the heater power to 5 or 10 % and rather permit
higher values if needed. If you plan to go to temperatures exceeding 100C, you
should possibly edit the regulation mode and use slope limited. That increases
temperature more gradually and avoids that the heater overheats.

Jochem Struppe



> From: main_at_ammrl.groups.io On Behalf Of Thusitha Jayasundera via groups.io
> Sent: Thursday, October 3, 2024 9:33 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Limit access to Temp. Display Bruker #Bruker #NMR_Aquisition_Software #NMR_Liquids

Hi Greg--did you mean limiting the accessible range by users? The admin can
do that via the edte window, under configuration; screencap attached.


Thusitha Jayasundera, Ph.D.
Director, Magnetic Resonance Center
Boston College
Merkert 105
2609 Beacon Street
Chestnut Hill MA 02467-3860


Tel: (617) 552-6984 (office)
Fax: (617) 552-2705<tel:617-552-2705>

Website: https://www.bc.edu/content/bc-web/research/sites/vice-provost-for-research/about/research-facilities/nmr-center


On Thu, Oct 3, 2024 at 9:24 AM gregory.wylie via groups.io wrote:

Hello All

We are running into an issue with users "training" other users to run variable
temperature on our new NEO 500. This is currently the only system we allow VT
on because we lost 3 older Varian probes last year during VT work. On the Varians
we had the temperature panel limited by having the software look at file on
startup. Does anyone have or know of away on to do something similar in Bruker
Software? I will admit we have not even tried yet, wanted to see if anyone was
doing it first.

Thanks
Greg

Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX 77843-3255
gpwylie_at_tamu.edu<mailto:gpwylie_at_tamu.edu>
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1701): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1701__;!!PvDODwlR4mBZyAb0!VNXOSlEyzuBWtCoRGcGa5S-XpPRmsC=
ZLu4xnPO2UMXSWuafIOYMgXEc6is-NhB95GOMeGKnXqhqjMXPFpg7YPMAGiihf$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108798757=
/7559972__;!!PvDODwlR4mBZyAb0!VNXOSlEyzuBWtCoRGcGa5S-XpPRmsCZLu4xnPO2UMXSWu=
afIOYMgXEc6is-NhB95GOMeGKnXqhqjMXPFpg7YPEombUTC$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!VNXOSlEyzuBWtCoRGcGa5S-XpPRmsCZLu4xnPO=
2UMXSWuafIOYMgXEc6is-NhB95GOMeGKnXqhqjMXPFpg7YPJM0CJ-3$
Mute #nmr_aquisition_software:https://urldefense.com/v3/__https://ammrl.gro=
ups.io/g/main/mutehashtag/nmr_aquisition_software__;!!PvDODwlR4mBZyAb0!VNXO=
SlEyzuBWtCoRGcGa5S-XpPRmsCZLu4xnPO2UMXSWuafIOYMgXEc6is-NhB95GOMeGKnXqhqjMXP=
Fpg7YPNOjGzXQ$
Mute #nmr_liquids:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!VNXOSlEyzuBWtCoRGcGa5S-XpPRm=
sCZLu4xnPO2UMXSWuafIOYMgXEc6is-NhB95GOMeGKnXqhqjMXPFpg7YPNHHvoOk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Oct 03 2024 - 06:51:43 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST