Re: [AMMRL] Limit access to Temp. Display Bruker #Bruker #NMR_Aquisition_Software #NMR_Liquids

From: Perry Pellechia <pellechia_at_sc.edu>
Date: Thu, 3 Oct 2024 13:40:32 +0000

Grep,

In the VT Control suite (at least for TS 3.x) under the Configuration tab you
should find a “set" button under the Temperature limits. Here you
can set a range smaller than the probe specified values. For extra safety
you can limit the Maximum power. I believe TS 4.x should have similar controls.

Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry



On Oct 3, 2024, at 9:21 AM, gregory.wylie via groups.io wrote:

Hello All

We are running into an issue with users "training" other users to run variable
temperature on our new NEO 500. This is currently the only system we allow VT
on because we lost 3 older Varian probes last year during VT work. On the
Varians we had the temperature panel limited by having the software look at
file on startup. Does anyone have or know of away on to do something similar
in Bruker Software? I will admit we have not even tried yet, wanted to see
if anyone was doing it first.

Thanks
Greg

Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX 77843-3255
gpwylie_at_tamu.edu<mailto:gpwylie_at_tamu.edu>
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1700): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1700__;!!PvDODwlR4mBZyAb0!Tuv8mL2pFJxiGhe9GfpifrsLvWWW8a=
b6CRaZpT65iNJCsOrspONTLXoZgtvwkLF_i8VGV5MWkG8Fgzwt9A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108798757=
/7559972__;!!PvDODwlR4mBZyAb0!Tuv8mL2pFJxiGhe9GfpifrsLvWWW8ab6CRaZpT65iNJCs=
OrspONTLXoZgtvwkLF_i8VGV5MWkG_IAVnYrw$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Tuv8mL2pFJxiGhe9GfpifrsLvWWW8ab6CRaZpT=
65iNJCsOrspONTLXoZgtvwkLF_i8VGV5MWkG8f2gbwyQ$
Mute #nmr_aquisition_software:https://urldefense.com/v3/__https://ammrl.gro=
ups.io/g/main/mutehashtag/nmr_aquisition_software__;!!PvDODwlR4mBZyAb0!Tuv8=
mL2pFJxiGhe9GfpifrsLvWWW8ab6CRaZpT65iNJCsOrspONTLXoZgtvwkLF_i8VGV5MWkG9IsdW=
gPw$
Mute #nmr_liquids:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!Tuv8mL2pFJxiGhe9GfpifrsLvWWW=
8ab6CRaZpT65iNJCsOrspONTLXoZgtvwkLF_i8VGV5MWkG_hsJzY-g$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






PastedGraphic-1.png
(image/png attachment: PastedGraphic-1.png)

Received on Thu Oct 03 2024 - 06:51:27 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST