On Oct 1, 2024, at 2:55 PM, gregory.wylie via groups.io wrote:
Hi EmilyIt actually maybe that the probe is not seated with the upper barrel well. This is easier with 2 people. But loosen the probe collar with the Allen wrench, and push up on the probe. This will push the upper barrel up. Then push the upper barrel back down. This should seat them together well. Retighten the probe collar.This is the main cause I have seen over the years.Greg_._,_._,_
Groups.io Links:You receive all messages sent to this group.
View/Reply Online (#1= 685) | Reply To= Group | Reply To Sender | Mute This Topic | New Topic
Mute #varian
Your Subscription | Contact Group Owner |_._,_._,_--Apple-Mail-EB926E2D-3326-4EB1-81AF-1A7CAE96FD37-- Received on Tue Oct 01 2024 - 16:47:06 MST