Re: [AMMRL] Error Lights on Lock Board, VNMRS (DDR1) #NMR_Liquids #Question #Troubleshooting #varian #VarianAgilent

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Mon, 11 Nov 2024 14:38:50 +0100

Dear Scott,

measuring 3.3 and 5.0 V is considerably too low - you should set the voltages
to something like 3.4-3.5 V and 5.2 V at least, otherwise not enough voltage
is "left" to arrive at the chip level after all the cables and plugs.
Please also run "Lock Scan" and watch out for a glitch at the first data point
(in real or imaginary part) - this could be an indication for a failing mixer
chip on the lock XCVR board.

Cheers
Bert

> Am 11.11.2024 um 04:20 schrieb Scott Burt via groups.io :
>
> Our 500 MHz VNMRS (DDR1) system has been having intermittent lock channel
> issues for a month or so. It really only surfaced as shimming failures and
> only happened with one specific user who likes to use mixed solvents. Every
> time this problem occurred, I ran some basic tests (lineshape, gradient profile,
> etc.) and everything worked fine, so I was convinced the problem was their
> samples. However, we started getting errors on most samples this past Thursday
> and then on Friday the lock control board started giving an error pattern in
> the blinking yellow lights.
>
> Error signal on the lock control board:
> The LEDs alternate between two states:
> only DS2 + DS5 illuminated
> only DS4 illuminated
>
> Is there a list somewhere of what the various error codes are for these boards?
> The other boards all have the normal "running" LEDs.
>
> Power cycling doesn't change anything.
> I've done some simple checking of the connections on the lock path.
>
> There's no obvious problem with the power supplies.
> Console power supply:
> measured (expected)
> +3.3 (3.3)
>
> +5.0 (5.0)
>
> +7.7 (7.6)
>
> +10.9 (10.5)
>
> -10.7 (-10.5)
>
> +15.2 (15.0)
>
> -15.2 (-15.0)
>
> +24.1 (24.0)
>
> RF front end:
> measured (expected)
> 18.0 (18.0)
>
> 7.0 (7.0)
>
> -18.0 (-18)
>
>
>0
> Technically, the error lights on the lock control board first showed up about
> 24 hours after upgrading from OVJ 3.1 to OVJ 3.2. However, the sudden increase
> in problems on Thursday happened before the upgrade of OVJ and, combined with
> the intermittent shim errors we've seen over the past month or two, this looks
> like a hardware error unrelated to the recent software update.
>
> Any suggestions on what to try next?
>
> Thanks!
>
> -Scott
>
> Dr. Scott Burt
> Teaching Professor
> NMR Lab Manager
> Department of Chemistry and Biochemistry
> Brigham Young University
> C414 BNSN / C008A BNSN
> Provo, Utah 84602-5700
> Phone: (801) 422-2404
> website: https://chembio.byu.edu/nmr-facility
> email: scott.burt_at_byu.edu
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1863): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1863__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL1edH9O=
Q7qNBGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE_WIr43qA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109512673=
/7559972__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL1edH9OQ7qNBGp9hU1jw2G=
9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE-SlM_pnQ$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL=
1edH9OQ7qNBGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE-t6n14oA$
Mute #nmr_liquids:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL1edH=
9OQ7qNBGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE_aNnOOHA$
Mute #question:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/question__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL1edH9OQ7qN=
BGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE9H7gwb9g$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmA=
MGfL1edH9OQ7qNBGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE9CwGL_Ww$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKePmAMGfL1edH9OQ7qNBGp9=
hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE-fBxGSYg$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!SMAzE-HN2WYZBpEKe=
PmAMGfL1edH9OQ7qNBGp9hU1jw2G9_HARFd6KP7P1QGDG8z9Be7S7eWhFHVtE_fbCIynw$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-



Received on Mon Nov 11 2024 - 06:25:54 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST