Re: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down

From: Gu, Jun via groups.io <jungu=sas.upenn.edu_at_groups.io>
Date: Fri, 8 Nov 2024 20:05:42 +0000

Hello All,

I would like to thank you for your kind response, with special thanks to Marc
Horst, Simon Sham, Rick Fitch, Vernon Russell, Jeff Walton, Peter Thompson,
Martha Morton, Al Brown and K. Klika.

Here is a summary of the actions taken:


1. I purchased three 230L high-pressure liquid nitrogen tanks. Thank you,
Martha and Peter, for providing valuable consumption data.
2. One high-pressure LN2 tank supplied gas to a 400 with SampleCase and
BCU I (400 l/h), a 500 with BCU II (400 l/h), and a 500 with a helium-cooled
cryoprobe (800 l/h). The LN2 tank was swapped after two days. The T connection
idea was extremely helpful!
3. Another high-pressure LN2 tank supplied gas to a 600 with a SampleCase,
BCU I (535 l/h) and a prodigy probe throughout the entire shutdown period
(four days), with the tank still half-full at the end.

Thank you for suggesting the VT turnoff. I chose to keep the VT on to understand
LN2 consumption without any spectrometer setting changes for future reference,
which explains why the tank supplying three NMRs lasted only two days. If I had
an extra regulator, I would have used two separate LN2 Dewars to support these
three NMRs.

Here are my recommendation:

1. Both cryoprobes use nitrogen gas cylinders for backup pressure, so swapping
tanks doesn’t risk probe warm-up.
2. Before connecting to the NMR, ensure the LN2 tank pressure is sufficiently
built up. Check the pressure in the morning and afternoon to keep it at least
above 120 psi.
3. Keep in mind that the gauge reading may not be accurate. Check the remaining
LN2 by shaking the tank and swap it before the pressure drops.
4. Purchase an extra high-pressure liquid nitrogen tank as a backup

I hope this information is helpful.

Best,

Jun

Jun Gu, PhD
Director of NMR Facility
Department of Chemistry
University of Pennsylvania
231 South 34th Street
Philadelphia, PA, 19104-6323
Phone: 215-898-1302
Office: Chem1973 Rm349

> From: main_at_ammrl.groups.io On Behalf Of Gu, Jun
> Sent: Wednesday, October 16, 2024 1:54 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down

Hello All,

There will be a house nitrogen shutdown for 3 day in my building. To keep the
NMRs available for research, I am considering using high-pressure liquid
nitrogen tanks to supply three spectrometers with SampleCase: (1)NMR with
a IProbe and BCU I, (2) NMR with a Prodigy probe and BCU I, (3) NMR with a
Helium-cooled probe (no SampleCase).

Is this approach feasible? How many liquid nitrogen tank should I order?
I would greatly appreciate it if anyone could share their experience or advice.

Best,

Jun


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1860): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1860__;!!PvDODwlR4mBZyAb0!TBx6jk4Lj4f5IWYsuCDOGwU2zlzQpE=
rubNboB0ihkY0HXP3bbqQDeQO0DVSDFSkc9nsdrI0T_vDRXdV5p9YVnjkVtgEP$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109048916=
/7559972__;!!PvDODwlR4mBZyAb0!TBx6jk4Lj4f5IWYsuCDOGwU2zlzQpErubNboB0ihkY0HX=
P3bbqQDeQO0DVSDFSkc9nsdrI0T_vDRXdV5p9YVnuJNZpCQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Nov 08 2024 - 12:08:48 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST