Re: [AMMRL] 1H and 13C NMR of Uranium (VI) complexes

From: Heiden, Zachariah M via groups.io <zachariah.heiden=wsu.edu_at_groups.io>
Date: Wed, 6 Nov 2024 16:41:07 +0000

Dear Ken,

We have users that daily measure radioactive samples (primarily U and Th
containing samples) at our instrumentation facility. To help with the fear
that regular users have with other users analyzing radioactive samples in
the same instrument facility we have strict operating procedures for the
analysis of radioactive samples (special designated locations and spinners
for radioactive samples, only manual sample insertion, double containment
of samples, post-analysis swipes and surveys, etc.). You can see our operating
procedures here:
https://nsc.wsu.edu/radioactive-sample-analysis/
Users that analyze radioactive NMR samples are required to place their samples
in a PTFE NMR tube liner (Wilmad Item #6005), sealed with a PTFE plug and placed
in a 5 mm NMR tube. The PTFE liner is very good containment if the glass NMR tube
breaks to prevent contamination of an area or an instrument.

Please let me know if you have any further questions.

Thank you,
Zach

Zachariah Heiden
ASSOCIATE DIRECTOR
Nuclear and Chemical Science (NUCS) Core Facility &
Nuclear Science Center
Washington State University
Direct Office: 509-335-0936 | Front Office: 509-335-8641
Email: zachariah.heiden_at_wsu.edu
https://nsc.wsu.edu/




> From: main_at_ammrl.groups.io On Behalf Of Kenneth Sharp-Knott
> Sent: Tuesday, November 5, 2024 9:30 AM
> To: ammrl_at_groups.io
> Subject: [AMMRL] 1H and 13C NMR of Uranium (VI) complexes


Hi Everyone,

I've had a request to perform 1H and 13C NMR on Uranium (VI) containing
complexes. The question is if it is safe or a hazard to the instrument.
I've been provided papers where it is done in relatively standard NMR setups,
and no mention is made of the fact that Uranium is present in the experimental
NMR section.

Does anyone have experience with this?

Best

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1852): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1852__;!!PvDODwlR4mBZyAb0!TafXt64X_1tgfaPt92jw_hAnya907i=
IbijVE7eKhvinAI-VVLX7-4wa7T4pdA5oL3aVHROchInH7WRskG1xp085TMmGN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109410390=
/7559972__;!!PvDODwlR4mBZyAb0!TafXt64X_1tgfaPt92jw_hAnya907iIbijVE7eKhvinAI=
-VVLX7-4wa7T4pdA5oL3aVHROchInH7WRskG1xp01NRYHoA$
Group Owner: main+owner_at_ammrl.groups.io



Received on Wed Nov 06 2024 - 08:46:59 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST