Re: [AMMRL] B-ACS sample in Shim error

From: Sebastian Kemper via groups.io <sebastian.kemper=tu-berlin.de_at_groups.io>
Date: Wed, 6 Nov 2024 14:20:37 +0000

Hallo Andre,

war die Fehlermeldung wirklich bei euch "15 SHIM SYSTEM NOT EMPTY”
oder „sample detect at magnet failed“?
Meiner Meinung nach müsste "15 SHIM SYSTEM NOT EMPTY” ein Fehler
des sample down sensors sein und
„sample detect at magnet failed“ ein Fehler des Sample Up
sensors.

Viele Grüße,
Sebastian


-----Original Message-----
> From: main_at_ammrl.groups.io On Behalf Of Andre Dallmann
> Sent: Dienstag, 21. November 2023 16:13
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] B-ACS sample in Shim error

Hi Pedro,  we do have that error from time to time, but if you take a
cotton swab and clean the sensor in the upper shim stack (the sample up
sensor) with ethanol it should go away.

Cheers, André

Am 21.11.2023 um 15:46 schrieb Pedro Aguiar:
> Hello,
>
> We have recently been suffering from an error on our AVANCE II system
> equipped with a B-ACS 60 and with black “sample down unit” at
> the top of BST.
>
> ERROR: "15 SHIM SYSTEM NOT EMPTY”
>
> This error pops up occasionally when going to insert the next sample. Completed
> sample is removed from BST and placed in carousel. Carousel moves to next
> tube and the arm obtains the next tube and moves it over and down into BST.
> However, before the pincers open, the error occasionally occurs. No obvious
> rhyme or reason that I can tell.
>
> I have been present when it fails and the “sample down” LED
> on the sample down unit does not come on before the error.
>
> I suspect one of two sensors, the sample down unit itself and/or one of the
> various cables may be at fault.
>
>
> Question 1) Have others had similar issues in the past? If so, any hints
> on which part(s) to try test first?
>
>
> Question 2) Have people removed the Sample down unit altogether (living
> dangerously…)? I tried and ICON was unhappy (never registered as having
> loaded and would not move on). Did I perhaps miss an option in ICON? Need
> to uninstall Firmware (update of firmware mentioned in sample down unit
> install manual)?
>
> Thanks,
> Pedro
>
>
>
> ****************************************************************
> Pedro M. Aguiar
>
> Coordonnateur - Laboratoire de RMN
> Département de Chimie
> Complexe de Sciences, Campus MIL B-3309 L’Université de Montréal
> Montréal, QC H2V 0B3 Canada
>
> Email: pedro.aguiar_at_umontreal.ca
> Tel: 514-343-6111 ext. 31006
>
> http://chimie.umontreal.ca/en/accueil/
> https://sites.google.com/site/pedroaguiarresearch/
> *****************************************************************
>
> Je comprends que vos heures de travail ne soient peut-être pas identiques à
> mes heures de travail.
>
> S'il vous plaît répondre pendant vos heures de travail.
>
>
> I appreciate that your working hours may not be the same as my working hours.
>
> Please reply during your working hours.
>
>
--
Dr. André Dallmann
Leiter der NMR-Abteilung

Email: andre.dallmann_at_chemie.hu-berlin.de
Raum: 0'103
Telefon: 7341
Handy: 01797657855
ORCID-ID: https://orcid.org/0000-0002-0141-2635
Webseite: https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en

Adresse:
Humboldt Universität zu Berlin
Institut für Chemie
AG NMR
Brook-Taylor-Str. 2
12489 Berlin
Deutschland


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1849): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1849__;!!PvDODwlR4mBZyAb0!S1CSsARa-o68SNDITLYVh4eZxDNwnh=
okfFC22K4fzqeWG_fy123aqDhjwHKtPnaZoEeMUqtcUzfQJuLW7S0JeRKFnKLI$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102729624=
/7559972__;!!PvDODwlR4mBZyAb0!S1CSsARa-o68SNDITLYVh4eZxDNwnhokfFC22K4fzqeWG=
_fy123aqDhjwHKtPnaZoEeMUqtcUzfQJuLW7S0JeaZsyvYI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Wed Nov 06 2024 - 06:20:52 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST