Re: [AMMRL] Varian 400MR Failed to Insert Sample

From: Kenneth Sharp-Knott <kknott_at_vt.edu>
Date: Fri, 1 Nov 2024 14:11:54 +0000

Is this the 12 pack carousel?

You should see a little light on the carousel box light up if it's 'detecting'
the sample. If it's inserting the sample but not detecting it, then are a
few reason why that could be

  1. Dirty sensor in the bore (at the bottom)
  2. The potentiometer might need to be adjusted


Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-