[AMMRL] Varian 400MR Failed to Insert Sample

From: Stephen Nolte <Stephen.Nolte_at_millersville.edu>
Date: Fri, 1 Nov 2024 13:54:12 +0000


Hello,

Our Varian 400MR running VnmrJ 4.2 has been having intermittent errors. After
the study is submitted, the autosampler successfully retrieves the ejected
placeholder sample, returns it to the tray, picks up the new sample, and
inserts the new sample into the magnet. At this point, before the ProTune
window appears, the tray position on the screen turns red and the analysis
stops. Right clicking on the tray position and selecting Study Event List
shows the attached document with the error “Failed to insert sample into
the magnet,” even though the sample has certainly been inserted.

Sometimes choosing the Tools > Sample in Magnet… function to eject the
errored sample and return the placeholder sample to the magnet, and then
choosing Automation > New Automation [autodir] > New Automation Run, and
starting over with a new study is enough to overcome the error. Other times it’s
necessary to close VnmrJ completely and use the su acqproc command in the
linux terminal to stop and restart acquisition processes to overcome the
error.

Does anybody have any advice?

Thank you!
_______________________________
Stephen Nolte | Chemistry Technician
Millersville University
717-871-7413


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1809): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1809__;!!PvDODwlR4mBZyAb0!Q3AnfPdz3ygnTvirQA0b0ZDbjt4ciS=
uGFLhCvnW4fldtn4hNIWlt2-DCNGX_wvtNYJwxzBkEfVhjq2FR_0nliBrIhSyZ6Q$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109334297=
/7559972__;!!PvDODwlR4mBZyAb0!Q3AnfPdz3ygnTvirQA0b0ZDbjt4ciSuGFLhCvnW4fldtn=
4hNIWlt2-DCNGX_wvtNYJwxzBkEfVhjq2FR_0nliBoHetYURQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






From owner-ammrl_at_ammrl.groups.io Fri, 1 Nov 2024 07:05:24
Received: by 2002:a17:504:dee5:b0:1bfe:7eec:3cb0 with SMTP id kl5csp469378njb;
        Fri, 1 Nov 2024 07:05:24 -0700 (PDT)
X-Forwarded-Encrypted: i=2; AJvYcCXHTHARqaj7QdKVJ/Hnkd5byvN/ikp7EYCrAx+ybw8gyczahQvvLt4YB+aAGJm0LJF1HgRn_at_hawaii.edu
X-Google-Smtp-Source: AGHT+IGz4FKKexpBAnoG3lHEqCM1gGpYXan9/dsEWzhdzMjzhm5mbSo+eDOn6iKMvt8HZEuavBWB
X-Received: by 2002:a05:6102:3707:b0:4a4:79cf:be83 with SMTP id ada2fe7eead31-4a9542b4691mr8130181137.10.1730469924312;
        Fri, 01 Nov 2024 07:05:24 -0700 (PDT)
ARC-Seal: i=1; a=rsa-sha256; t=1730469924; cv=none;
        d=google.com; s=arc-20240605;
        b=dmxVBlBvaed/DqreP8ndzEauQJEx//rlEaf68H5crjOBII7QS3xepw1w2Ut2I7YZS1
         f4h0qEmcK2A8OR1FDaOmNDLb50G89Wm18c3c6fb/4HQJY1I4Zc6A3AQU4zP8fm8ElHI6
         pBF2FlZH5SyfdIFcmd7A81H3d9nNWMeGsSGhfTlUIYDJZd3X+YIpW80NIzWpi+LFOcvk
         g+XukL2qtCR91oe0w9+IDnQq+pRMbxrwUnwDhOEfgOqmTtg+UJdm9/fwE3azM24XceXA
         tkEC7OUTjUQmMhbtk10NjZD5uIv0JgVZ5Al1xSlUyHNZVrMvStL2bWd1HFaKYKklk5DF
         SpXA==
ARC-Message-Signature: i=1; a=rsa-sha256; c=relaxed/relaxed; d=google.com; s=arc-20240605;
        h=mime-version:list-unsubscribe:list-unsubscribe-post:reply-to
         :resent-from:resent-date:delivered-to:mailing-list:list-id:sender
         :list-help:list-subscribe:precedence:to:subject:message-id:date:from
         :in-reply-to:references:dkim-signature:resent-message-id;
        bh=NBuvxQvRx28KHzWRTgMkEi6oPMRqzqYJhNUHBbTYFBY=;
        fh=xC55mW4WXjmdiwtkPJUr4APdxMPTmW+ubeamlxP9uX0=;
        b=F+1axcLAhW7Tf1NyqZAjCVDlHSOcewhGZaMS44H2KJQkvDfkzJwd2aZWQK3Olf9cGS
         7yukEYy5nT43nd2aoib80vuwCSA0EIHyL9I/cN1k5XaAzWY2BrreF95qU68SIB/6RIdM
         Ukz2cv/Y4v/o7ZaAJjRBvcy0hDa24UmmzZGNRDDO24Mw6RQnAJ0CJ8cjfKcXhzqgNXHr
         SNIAVoWCbFYNvfjBevcYonmKYrwDjjcugDFQo5SUJQ/Gp/x8rtJtIXneCvIZyldPhayO
         r+libWXVqLAmBEeEWxWGWFQKMnbqPRGh4ddh5P4ooix2BWdnYJ5JIu7gF9YabDiAKnl0
         fSzg==;
        dara=google.com
ARC-Authentication-Results: i=1; mx.google.com;
       dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=20240830 header.b=kR1XpLW5;
       spf=pass (google.com: domain of bounce+115296+1807+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) smtp.mailfrom=bounce+115296+1807+7559972+12177410_at_groups.io
Return-Path: <bounce+115296+1807+7559972+12177410_at_groups.io>
Received: from mx0a-007a6701.pphosted.com (mx0a-007a6701.pphosted.com. [205.220.169.170])
        by mx.google.com with ESMTPS id ada2fe7eead31-4a957489a29si852518137.198.2024.11.01.07.05.23
        for <walt_at_hawaii.edu>
        (version=TLS1_2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128/128);
        Fri, 01 Nov 2024 07:05:24 -0700 (PDT)
Received-SPF: pass (google.com: domain of bounce+115296+1807+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) client-ip=192.53.124.254;
Authentication-Results: mx.google.com;
       dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=20240830 header.b=kR1XpLW5;
       spf=pass (google.com: domain of bounce+115296+1807+7559972+12177410_at_groups.io designates 192.53.124.254 as permitted sender) smtp.mailfrom=bounce+115296+1807+7559972+12177410_at_groups.io
Received: from pps.filterd (m0313020.ppops.net [127.0.0.1])
        by mx0a-007a6701.pphosted.com (8.18.1.2/8.18.1.2) with ESMTP id 4A12fnK6028034
        for <walt_at_hawaii.edu>; Fri, 1 Nov 2024 04:05:23 -1000
Resent-Message-Id: <202411011405.4A12fnK6028034_at_mx0a-007a6701.pphosted.com>
Authentication-Results: ppops.net;
        spf=pass smtp.mailfrom=bounce+115296+1807+7559972+12177410_at_groups.io;
        dkim=pass header.s=20240830 header.d=groups.io;
        dmarc=fail header.from=gmail.com
Received: from mail08.groups.io (mail08.groups.io [192.53.124.254])
        by mx0a-007a6701.pphosted.com (PPS) with ESMTPS id 42mpe8skht-1
        (version=TLSv1.2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128 verify=NOT)
        for <walt_at_hawaii.edu>; Fri, 01 Nov 2024 04:05:22 -1000 (HST)
DKIM-Signature: a=rsa-sha256; bh=flXZNQpr5bK55WP5D5je8QdRHyU7WhTDGpvt6+wSYko=;
 c=relaxed/simple; d=groups.io;
 h=MIME-Version:References:In-Reply-To:From:Date:Message-ID:Subject:To:Precedence:List-Subscribe:List-Help:Sender:List-Id:Mailing-List:Delivered-To:Resent-Date:Resent-From:Reply-To:List-Unsubscribe-Post:List-Unsubscribe:Content-Type;
 s=20240830; t=1730469922; v=1; x=1730729122;
 b=kR1XpLW5JKO5n4wf/dJFe+QuTY+GmlZLUpEhnxB0flggyMOlV1mArgMRE178QvAl9gvnU9cy
 wrKPAVajHkewmp/qDir+wPoWH64m6iph/bQht8SSB6Xr8bVsS2SOPcw6+EdYBc1jhjGSaCNRMTJ
 yReUUPn7A8KMUeyFMQ3As0FruUxCZe/n5Hhxnmt3iFRI9aORnBc00rdjoFJ24roGuyArS0M8Bq+
 EtTk3QAhhmVhTl8kDyQpG/SS/mqWAhm5Flq9L/N3sCyGkdFfFEN/IIyWEa2GSJ/S3/at61m9/3Q
 AiQapj/lBZqgGlMOtnZTJogYoGRyoTocLk8t84zfyTKzQ==
X-Received: by 127.0.0.2 with SMTP id ytJhYY7561307xL45THhiERi; Fri, 01 Nov 2024 07:05:22 -0700
X-Received: from mail-lj1-f180.google.com (mail-lj1-f180.google.com [209.85.208.180])
 by mx.groups.io with SMTP id smtpd.web11.38426.1730469916611984876
 for <main_at_ammrl.groups.io>;
 Fri, 01 Nov 2024 07:05:16 -0700
X-Received: by mail-lj1-f180.google.com with SMTP id 38308e7fff4ca-2fb4fa17044so21865881fa.3
        for <main_at_ammrl.groups.io>; Fri, 01 Nov 2024 07:05:16 -0700 (PDT)
X-Gm-Message-State: M0SYgIlNPgTuROzoLHhNgF9vx7559972AA=
X-Received: by 2002:a2e:a90a:0:b0:2fa:ce00:3689 with SMTP id
 38308e7fff4ca-2fedb75731fmr23792251fa.6.1730469913416; Fri, 01 Nov 2024
 07:05:13 -0700 (PDT)
References: <6de1e8b38c5949f7b5bf28bdd51358f5_at_exxonmobil.com> <BL0PR01MB5283ED9000038FBAAC407AF5DA562_at_BL0PR01MB5283.prod.exchangelabs.com>
In-Reply-To: <BL0PR01MB5283ED9000038FBAAC407AF5DA562_at_BL0PR01MB5283.prod.exchangelabs.com>
From: "Craig Grimmer" <craig.grimmer_at_gmail.com>
Date: Fri, 1 Nov 2024 15:05:00 +0100
Message-ID: <CAP9hRSu8Y-HOcGV=u1Ljz9FQo1z-s3EZWjZO2zKyuVtJ7nDt3w_at_mail.gmail.com>
Subject: Re: [AMMRL] Need help converting Bruker neo data into Varian
To: main_at_ammrl.groups.io
Precedence: Bulk
List-Subscribe: <mailto:main+subscribe_at_ammrl.groups.io>
List-Help: <mailto:main+help_at_ammrl.groups.io>
Sender: main_at_ammrl.groups.io
List-Id: <main.ammrl.groups.io>
Mailing-List: list main_at_ammrl.groups.io; contact main+owner_at_ammrl.groups.io
Delivered-To: mailing list main_at_ammrl.groups.io
Resent-Date: Fri, 01 Nov 2024 07:05:17 -0700
Resent-From: craig.grimmer_at_gmail.com
Reply-To: main_at_ammrl.groups.io
List-Unsubscribe-Post: List-Unsubscribe=One-Click
List-Unsubscribe: <https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/plugh>
Content-Type: multipart/alternative; boundary="0000000000007bdd8f0625da6cc8"
X-Proofpoint-ORIG-GUID: AmXiWP6khYZhhnTeVtUfw18An77o-q3a
X-CLX-Shades: MLX
X-Proofpoint-GUID: AmXiWP6khYZhhnTeVtUfw18An77o-q3a
X-CLX-Response: 1TFkXGxkbEQpMehcYEx4RCllEF2JeeUVJSRhDUGhJEQpYWBdubl1TGmdCbmR 7bhEKeE4XaGxnQRJLaXIYfXMRCnhLF25uXVMaZ0JuZHtuEQp5TBdhSx18UEhTXkAZRxEKQ0gXBx gZHREKQ1kXBxsaGREKQ0kXGgQaGhoRCllNF2dmchEKX1kXGx0eEQpfTRdnZnIRCllJFx8ScRsbB
 hsadwYbHRMGGgYHHRpCGB0GEx0GGxsbcRsaGBAddwYaBgcYGhoGGgYaBhoGGnEaEBp3BhoRClle F2hueREKSUYXX0RDXE9YWUNeU0JLXUtDQ3VCRVleT04RCklHF3hPTREKQ04Xa0dyQ316HEFCc3B CQkR+T3xef0xdGxJrRB0dRQdbGUsRClhcFx8EGgQZHRMFGxoEGxsaBBsZHgQZGRAbHhofGhEKXl
 kXT31FeRMRCk1cFxsTHxEKTFoXaWxtXUBTEQpMRhdea2sRCkJPF20eUkcBGHBIZnJrEQpDWhcbE xgEHxkEGxgeBBgfHhEKQl4XGxEKQkYXYEVuQ0tHQ11rHEwRCkJHF2JdXhpwR3NOZXpBEQpCRRd6 WmN/Z01SeG9cEhEKQk4XY1Nja3sTWH4aX14RCkJMF2dNSR9EHgVzYxhDEQpCbBdpclJlWhlFY15
 cYREKQkAXYGFyeBN8eEVvaxMRCkJYF25waGNYe258G1NMEQpaWBccEQp5QxdiYXlPRGZge3gcfB EKWUsXGxocHRkRClpLFx8bGh0RCnBnF29uYUBARWFsemseEBoRCnBoF24FbBlYS3NNTWl+EBoRC nBoF21SZ2Z7GG4ZHmJnEBoRCnBoF2NNbU9cYWdpQX1nEBoRCnBoF2MYeltEewFtBW5EEBoRCnBo
 F2gTWhxsZF1pW0lwEBoRCnBoF21ZX2ABE1BdYxxiEAceGhEKcGgXa3tsGB9uaURjeVsQGhEKcGg XY0NMch9JYU5oW04QGhEKcH0XZ39tcGdkbl4Fbx4QBxkaEQpwZxdmAUceTx1iU2MfSRAHHhoRCn BnF20bHWlwfm1TRFJPEAcZGhEKcGcXZQVNbGFEAUBEa3AQBxkaEQpwfRdjfB1BH0Noc11rYRAaE
 QpwfRdmcgVwa0BfTGV/SBAaEQpwZxdlZllsXkdoH09ffRAHHhoRCnB9F2dsH1B/Q05GU0MBEBoR CnBnF2xsXUFOHU4dXk9lEBoRCnB9F29neVsZaUYaHE9PEAceHBEKcGcXYl9+Y0JEcGVCf2MQGhE KcH0XZUB7E1BAWhtaW2QQBxkaEQpwZxd6YkQbZkRfH0dfaBAaEQpwfRdseHweW1IaZV4cYhAHGR
 oRCnB9F2FuSAFac3xuGWFbEBoRCnB/F2cfQV1MSWxhY2hyEAcZGhEKcF8XbXNpXxx4RnAfYVMQB xkaEQpwfxdsGVkBbWQSZkUZfhAHHhoRCnBfF2ZaW1xvR0tuck5TEAccGhEKcF8XbFx7Em1vE09w H0UQBx4bEQpwXxdofE9gUG1oa1keYxAHGRoRCnB/F29neVsZaUYaHE9PEAceHBEKcF8XblB+Yx4
 ac31rZEEQBxkaEQpwXxdseHweW1IaZV4cYhAHGRoRCnBfF2sFGmh9XVBSYHATEAcZGhEKcGwXZk wbXkZNRR1acGYQGhEKbX4XGhEKWE0XSxEg
MIME-Version: 1.0
X-Proofpoint-Virus-Version: vendor=baseguard
 engine=ICAP:2.0.293,Aquarius:18.0.1039,Hydra:6.0.680,FMLib:17.12.60.29
 definitions=2024-09-06_09,2024-09-06_01,2024-09-02_01
X-Proofpoint-Spam-Details: rule=inbound_notspam policy=inbound score=0 spamscore=0 phishscore=0
 mlxlogscore=999 clxscore=131 mlxscore=0 bulkscore=0 suspectscore=0
 malwarescore=0 lowpriorityscore=0 adultscore=0 priorityscore=294
 impostorscore=0 unknownsenderscore=20 classifier=spam adjust=0 reason=mlx
 scancount=1 engine=8.19.0-2409260000 definitions=main-2411010102
 domainage_hfrom=10673 domainage_replyto=5107

--0000000000007bdd8f0625da6cc8
Content-Type: text/plain; charset="UTF-8"
Content-Transfer-Encoding: quoted-printable

Hi Sarah

I haven't tried this but you may be able to do it in two steps: firstly,
convert your Neo-format into AMX-format with "convdta" (in TopSpin);
secondly, read your AMX-format with "convertbru" (in VNMR).

Kind regards,
Craig.


On Fri, 1 Nov 2024 at 14:38, D'Anciaes Almeida Silva, Igor via groups.io wrote:

> Hi Sarah,
>
> I never did it, but I'd try using NMRglue and Python (
> https://www.nmrglue.com/
>
> Best,
> ----------------
> Dr. Igor d'Anci=C3=A3es Almeida Silva
> Research Scientist - NMR Core
> University of Missouri-Columbia
> 32A Chemistry Building
> ------------------------------
> *From:* main_at_ammrl.groups.io on behalf of Mattler,
> *Sent:* Thursday, October 31, 2024 4:40 PM
> *To:* main_at_ammrl.groups.io
> *Subject:* [AMMRL] Need help converting Bruker neo data into Varian
>
> Hi all,
>
> I am hoping I can get some help. I am having issues converting some Bruker
> Neo data into Varian. Has anyone tried this and been successful? Any help
> you can provide would be useful. Thanks!
>
> Sincerely,
>
>
> *Sarah Mattler, PhD*
>
> sarah.j.mattler_at_exxonmobil.com
>
>

-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1807): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1807__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsziN=
XTKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtLg1U97Ug$=20
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109323607=
/7559972__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsziNXTKTdhH569I4k5a=
gZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtI4aNCHAw$=20
Group Owner: main+owner_at_ammrl.groups.io
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-



--0000000000007bdd8f0625da6cc8
Content-Type: text/html; charset="utf-8"
Content-Transfer-Encoding: quoted-printable

<div dir=3D"ltr"><div>Hi Sarah</div><div><br/></div><div>I haven&#39;t trie=
d this but you may be able to do it in two steps: firstly, convert your Neo=
-format into AMX-format with &#34;convdta&#34; (in TopSpin); secondly, read=
 your AMX-format with &#34;convertbru&#34; (in VNMR).</div><div><br/></div>=
<div>Kind regards,</div><div>Craig.</div><div><br/></div></div><br/><div cl=
ass=3D"gmail_quote"><div dir=3D"ltr" class=3D"gmail_attr">On Fri, 1 Nov 202=
4 at 14:38, D&#39;Anciaes Almeida Silva, Igor via <a href=3D"https://urldef=
ense.com/v3/__http://groups.io__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8=
aXbsziNXTKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtJVRBHiQw$">groups.io=
</a> &lt;idhbx=3D<a href=3D"mailto:missouri.edu_at_groups.io">missouri.edu_at_gro=
ups.io</a>&gt; wrote:<br/></div><blockquote class=3D"gmail_quote" style=3D"=
margin:0px 0px 0px 0.8ex; border-left:1px solid rgb(204,204,204); padding-l=
eft:1ex;"><div class=3D"msg5647234918696999969">



<div dir=3D"ltr">
<div>
Hi Sarah,</div>
<div>
<br/>
</div>
<div>
I never did it, but I&#39;d try using NMRglue and Python (<a href=3D"https:=
//urldefense.com/v3/__https://www.nmrglue.com/__;!!PvDODwlR4mBZyAb0!XLIfJMG=
UwkxuCLa98dkrXo8aXbsziNXTKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtIirm=
G-aQ$" id=3D"m_5647234918696999969LPlnk500536" target=3D"_blank">https://ww=
w.nmrglue.com/</a>)</div>
<div id=3D"m_5647234918696999969Signature">
<div>
<br/>
</div>
<div>
Best,</div>
<div>
----------------</div>
<div>
Dr. Igor d&#39;Anci=C3=A3es Almeida Silva</div>
<div>
Research Scientist - NMR Core</div>
<div>
University of Missouri-Columbia</div>
<div>
32A Chemistry Building</div>
</div>
<div id=3D"m_5647234918696999969appendonsend"></div>
<hr style=3D"display:inline-block; width:98%;"/>
<div id=3D"m_5647234918696999969divRplyFwdMsg" dir=3D"ltr"><u></u><b>From:<=
/b> <a href=3D"mailto:main_at_ammrl.groups.io" target=3D"_blank">main_at_ammrl.gr=
oups.io</a> &lt;<a href=3D"mailto:main_at_ammrl.groups.io" target=3D"_blank">m=
ain_at_ammrl.groups.io</a>&gt; on behalf of Mattler, Sarah J via <a href=3D"ht=
tps://urldefense.com/v3/__http://groups.io__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkx=
uCLa98dkrXo8aXbsziNXTKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtJVRBHiQw=
$" target=3D"_blank">groups.io</a> &lt;sarah.j.mattler=3D<a href=3D"mailto:=
exxonmobil.com_at_groups.io" target=3D"_blank">exxonmobil.com_at_groups.io</a>&gt=
;<br/>
<b>Sent:</b> Thursday, October 31, 2024 4:40 PM<br/>
<b>To:</b> <a href=3D"mailto:main_at_ammrl.groups.io" target=3D"_blank">main_at_a=
mmrl.groups.io</a> &lt;<a href=3D"mailto:main_at_ammrl.groups.io" target=3D"_b=
lank">main_at_ammrl.groups.io</a>&gt;<br/>
<b>Subject:</b> [AMMRL] Need help converting Bruker neo data into Varian<u>=
</u>
<div>=C2=A0</div>
</div>

<div lang=3D"EN-US">
<div style=3D"width:100%; border-style:solid; border-color:rgb(156,0,6); bo=
rder-width:1pt; padding:2pt; line-height:12pt; text-align:left; margin-bott=
om:2pt;">
<span>WARNING:</span> This message has originated from an External Source. =
This may be a phishing expedition that can result in unauthorized access to=
 our IT System. Please use proper judgment and caution when opening
 attachments, clicking links, or responding to this email.</div>
<div>
<div>
<p>Hi all,</p>
<p>=C2=A0</p>
<p>I am hoping I can get some help. I am having issues converting some Bruk=
er Neo data into Varian. Has anyone tried this and been successful? Any hel=
p you can provide would be useful. Thanks!</p>
<p>=C2=A0</p>
<p><span>Sincerely,</span></p>
<p><span>=C2=A0</span></p>
<p><span>=C2=A0</span></p>
<p><b><span>Sarah Mattler, PhD</span></b></p>
<p><span><a href=3D"mailto:sarah.j.mattler_at_exxonmobil.com" target=3D"_blank=
">sarah.j.mattler_at_exxonmobil.com</a><b></b></span></p>
<p>=C2=A0</p>
</div>

</div>
</div>






<p></p><p></p></div></div></blockquote></div>


<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
<hr>


Groups.io Links:<p>


 =20
    You receive all messages sent to this group.
 =20
 =20


<p>
<a target=3D"_blank" href=3D"https://urldefense.com/v3/__https://ammrl.grou=
ps.io/g/main/message/1807__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsz=
iNXTKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtLg1U97Ug$">View/Reply Onl=
ine (#1807)</a> |


  <a target=3D"_blank" href=3D"mailto:main_at_ammrl.groups.io?subject=3DRe:%20=
Re%3A%20%5BAMMRL%5D%20Need%20help%20converting%20Bruker%20neo%20data%20into=
%20Varian">Reply to Group</a>
 =20
    | <a target=3D"_blank" href=3D"mailto:craig.grimmer_at_gmail.com?subject=
=3DPrivate:%20Re:%20Re%3A%20%5BAMMRL%5D%20Need%20help%20converting%20Bruker=
%20neo%20data%20into%20Varian">Reply to Sender</a>
 =20

|

  <a target=3D"_blank" href=3D"https://urldefense.com/v3/__https://groups.i=
o/mt/109323607/7559972__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsziNX=
TKTdhH569I4k5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtI4aNCHAw$">Mute This Topic</=
a>


| <a href=3D"https://urldefense.com/v3/__https://ammrl.groups.io/g/main/pos=
t__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsziNXTKTdhH569I4k5agZAdfxB=
c9E_d0LydGzkf1FE7D79KlHPFtJna6GvVQ$">New Topic</a>

<br>




<a href=3D"https://urldefense.com/v3/__https://ammrl.groups.io/g/main/edits=
ub/7559972__;!!PvDODwlR4mBZyAb0!XLIfJMGUwkxuCLa98dkrXo8aXbsziNXTKTdhH569I4k=
5agZAdfxBc9E_d0LydGzkf1FE7D79KlHPFtItiWZ3ZQ$">Your Subscription</a> |
<a href=3D"mailto:main+owner_at_ammrl.groups.io">Contact Group Owner</a> |

<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>


--0000000000007bdd8f0625da6cc8--

Received on Fri Nov 01 2024 - 07:05:54 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST