Hello Kenneth,
I echo Alex Patterson's comments.
- If it spins and tunes, you are fine. 5% doesn't strike me as too bad.
- Yes, Fe helps with T1, but broadens lines.
- A great concern is the uneven distribution, since it affects species
unevenly, hence can disturb quantification.
- CP - I found that doing CP really distorts the intensities. For Si I
typically use one pulse excitation with a small tip angle and pulse faster.
- Quartz, yes you will see a beautiful Q4, once! It has a really long T1.
In one project I took my data, then let the sample sit overnight, spun it
up and took one scan. (of course no ds). Beautiful Q4, which I did not see
the day before.
Feel free to reach out directly, if you have questions. I have done quite a
bit of cement and other siliceous materials.
Happy resonating,
Ulli
On Tue, May 19, 2026 at 4:25 PM Alex Paterson via groups.io wrote:
> Hi Ken,
>
> 1. I do not personally have strict limits on paramagnetic content; if
> it tunes and spins, it’s fine. I’d use all caution when spinning up, but if
> there are susceptibility effects I’d expect them to be obvious at first
> attempt.
>
> 2. High paramagnetic content will decrease T1, but also T2. Depending
> on the distribution of paramagnetic ions in the sample, this can sometimes
> lead to either horrific resolution, or sometimes to cases where peaks of
> interest are simply invisible (while other, further environments are
> unchanged). I do not know of a way to predict this a priori. The
> distribution of paramagnetic ions can also affect quantitation; if the
> paramagnetic species are phase-segregated, they will preferentially relax
> some environments and not others. Doping with paramagnetic ions to reduce
> T1 homogeneously is common in the study of oxide glasses, but there I
> expect paramagnetic ions to be more evenly distributed than in an aggregate.
>
> 3. I wouldn’t call a signal from within your sample a background,
> simply an unwanted peak. The difference is important as common
> background-suppression techniques in SSNMR (Hahn echoes, DEPTH) rely on
> differential nutation of signals originating from outside the rotor. As
> your quartz will be within the rotor, conventional methods will not work.
> Spectral editing via CPMAS is one possibility, but that will affect the
> reliability of your quantitation. Another possibility, particularly if the
> iron is not present in the quartz, is using T1 editing; *if* the
> quartz has a much longer relaxation time than the rest of your sample,
> using presaturation loops prior to the rest of your experiment should
> reliably suppress the quartz signal. This will of course mean you’re unable
> to quantify the quartz species, but that may be an acceptable compromise.
>
> 4. As for 29Si{1H} CPMAS, I have found tetrakis(trimethysilyl)silane
> to be an excellent setup standard for that experiment. Note that you may
> need to use contact times well beyond what you’d expect from 13C{1H} CPMAS,
> and the high paramagnetic content may make CPMAS challenging as well.
>
> Best of luck,
>
> Alex
>
> Alex Paterson
>
> The National Magnetic Resonance Facility at Madison (NMRFAM)
>
> University of Wisconsin-Madison
>
> alpaterson_at_wisc.edu
>
> *From:* main_at_ammrl.groups.io <main_at_ammrl.groups.io> *On Behalf Of *Kenneth
> Sharp-Knott via groups.io
> *Sent:* Tuesday, May 19, 2026 1:25 PM
> *To:* main_at_ammrl.groups.io
> *Subject:* [AMMRL] Advice on 29Si SSNMR of concrete samples –
> Paramagnetic iron limits and T1 relaxation
>
>
>
> Hi everyone,
>
> We are finally fully up and running with solid-state NMR capabilities here
> at Virginia Tech (running a Neo400 with a 4 mm MAS probe), and I have a
> student requesting to run 29Si MAS NMR on some 60-year-old concrete samples.
>
> They mentioned that the samples contain about 5% Fe2O3, as determined by
> XRF. I have very little experience with concrete samples in SSNMR and
> wanted to tap into the community's expertise regarding a few points:
>
> 1. Spinning Safety: My understanding is that an iron oxide content of
> around 10% is the threshold where magnetic susceptibility begins to put the
> instrument at risk of a rotor crash or stator damage, but that anything
> below that is generally safe to spin. Can anyone shed some light on this or
> share their safety cutoffs for paramagnetic/ferromagnetic samples?
>
> 2. Relaxation Times: I've also read that there may actually be a benefit
> to this Fe2O3 content—namely that the paramagnetic iron will reduce the
> normally agonizing 29Si T1 relaxation times, allowing for shorter recycle
> delays for quantitative direct-excitation experiments. Can anyone confirm
> if this provides a meaningful practical benefit for cementitious samples?
>
> Finally, since these are crushed *concrete* samples (aggregate included)
> rather than extracted cement pastes, I anticipate a massive Q4 quartz
> background. If anyone has tips on the best way to handle or filter this (we
> are planning to try 1H-29Si CP-MAS to selectively highlight the C-S-H gel,
> I would greatly appreciate it.
>
> Thank you in advance for your insights!
>
> Best regards,
>
> Ken Sharp-Knott
> Manager of Analytical Services and the NMR Facility
> Department of Chemistry
> Virginia Tech
>
> (540)267-6502 (Cell)
> (540)231-0885 (Office)
>
--
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way
PO BOX 15000
Halifax, NS B3H 4R2
e-mail: Ulli.Zwanziger_at_Dal.ca
Tel. 1 902 494 8085
FAX: 1 902 494 1310
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3476): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3476__;!!PvDODwlR4mBZyAb0!TvYVPWgO9jXI0ipqatwFMVqOt0qTr4=
Gz7_40jfsbezQV13UCSkoMvCi2hu4kSq_KlRDkwT9vwvgttQ1RQGa9Wl6mic26$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119395828=
/7559972__;!!PvDODwlR4mBZyAb0!TvYVPWgO9jXI0ipqatwFMVqOt0qTr4Gz7_40jfsbezQV1=
3UCSkoMvCi2hu4kSq_KlRDkwT9vwvgttQ1RQGa9Wm3t5w08$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed May 20 2026 - 13:14:06 MST