[AMMRL] Advice on 29Si SSNMR of concrete samples – Paramagnetic iron limits and T1 relaxation

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Tue, 19 May 2026 18:25:12 +0000

Hi everyone,

We are finally fully up and running with solid-state NMR capabilities here
at Virginia Tech (running a Neo400 with a 4 mm MAS probe), and I have a student
requesting to run 29Si MAS NMR on some 60-year-old concrete samples.

They mentioned that the samples contain about 5% Fe2O3, as determined by XRF.
I have very little experience with concrete samples in SSNMR and wanted to
tap into the community's expertise regarding a few points:

1. Spinning Safety: My understanding is that an iron oxide content of around
10% is the threshold where magnetic susceptibility begins to put the instrument
at risk of a rotor crash or stator damage, but that anything below that is
generally safe to spin. Can anyone shed some light on this or share their
safety cutoffs for paramagnetic/ferromagnetic samples?

2. Relaxation Times: I've also read that there may actually be a benefit to
this Fe2O3 content—namely that the paramagnetic iron will reduce the normally
agonizing 29Si T1 relaxation times, allowing for shorter recycle delays for
quantitative direct-excitation experiments. Can anyone confirm if this
provides a meaningful practical benefit for cementitious samples?

Finally, since these are crushed concrete samples (aggregate included) rather
than extracted cement pastes, I anticipate a massive Q4 quartz background.
If anyone has tips on the best way to handle or filter this (we are planning
to try 1H-29Si CP-MAS to selectively highlight the C-S-H gel), I would greatly
appreciate it.

Thank you in advance for your insights!

Best regards,

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3470): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3470__;!!PvDODwlR4mBZyAb0!RCkoAogx63NLO6QaS-wSvvHAObSbWO=
WsmXu9j-TDzppjJ5zTdvMnV9wzLmOHnKgZyxs3FfepB0IRypTIrLhuqMsp$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119395828=
/7559972__;!!PvDODwlR4mBZyAb0!RCkoAogx63NLO6QaS-wSvvHAObSbWOWsmXu9j-TDzppjJ=
5zTdvMnV9wzLmOHnKgZyxs3FfepB0IRypTIrH6jq87p$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 19 2026 - 11:25:43 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST