Re: [AMMRL] Bruker AVANCE III Blown fuses?

From: Banghao Chen via groups.io <chen=chem.fsu.edu_at_groups.io>
Date: Tue, 19 May 2026 13:30:51 +0000

Hi Casey,

I had the same issue on one of my AVIII single bay consoles, the problem was
resolved after taking the switch out and replaced the FUSE CIRBREA BIP THERMIC (16A)
with one ordered from Bruker. Check with the Bruker sales manager in your area.

Good Luck!

BHC


******************************
Banghao Chen
FSU-NMR Facility Director,
Distinguished University Scholar
Florida State University
Department of Chemistry and Biochemistry
102 Varsity Way (CSL) 2001 Levy Ave (ICRB)
Tallahassee, FL 32306 Tallahassee, FL 32310
chen_at_chem.fsu.edu<mailto:chen_at_chem.fsu.edu>
(850) 644-3334

________________________________
> From: main_at_ammrl.groups.io on behalf of Casey Simons via groups.io
> Sent: Monday, May 18, 2026 6:42 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Bruker AVANCE III Blown fuses?

Hey all, we just had a power outage which flipped everything off. New to me
is that our advance III consoles main power switch will not stay depressed.
My guess is a blown fuse somewhere in the console. Does anyone have experience
dealing with this?

C

Casey Simons
NMR Director
CAMCOR Nanofabrication and Imaging Facility
University of Oregon



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3469): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3469__;!!PvDODwlR4mBZyAb0!XwxdIYzkxxqaDm6lxiMzDfRHfReHjK=
OvMsaVVepShjGgxxNpaykvJF_k022KJ-rt0SwCxeGl58RBD_ZxEnoZzYyXLcNO$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119383646=
/7559972__;!!PvDODwlR4mBZyAb0!XwxdIYzkxxqaDm6lxiMzDfRHfReHjKOvMsaVVepShjGgx=
xNpaykvJF_k022KJ-rt0SwCxeGl58RBD_ZxEnoZzZ-SmdvN$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 19 2026 - 06:32:49 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST