Re: [AMMRL] Bruker AVANCE III Blown fuses?

From: Andrew Fowler via groups.io <afowler13=gmail.com_at_groups.io>
Date: Tue, 19 May 2026 06:46:24 -0400

Hi Casey,

In back at the bottom is the safety box which should have a breaker style
power switch on the left, a few small "bricks" going across, and ends with
the accessory outlets on the right. Check that the switch is up/on first
then the more narrow piece just right of the switch should be a fuse holder
with a standard glass mini fuse in it (I think there's one fuse, possibly
2). Check to see if that's blown.

Good luck,
Andrew

On Mon, May 18, 2026 at 8:27 PM Casey Simons via groups.io wrote:

> Hey all, we just had a power outage which flipped everything off. New to
> me is that our advance III consoles main power switch will not stay
> depressed. My guess is a blown fuse somewhere in the console. Does anyone
> have experience dealing with this?
>
> C
>
> Casey Simons
> NMR Director
> CAMCOR Nanofabrication and Imaging Facility
> University of Oregon
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3468): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3468__;!!PvDODwlR4mBZyAb0!U6NiQg3C8aQ_LoZVeGYWkAr98GoCPE=
SqQvaudxmbuYIs0AQ7vwjrXlci79BrblbiILS1cxCeoRlcQuXXUy-Eilw7ADKb$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119383646=
/7559972__;!!PvDODwlR4mBZyAb0!U6NiQg3C8aQ_LoZVeGYWkAr98GoCPESqQvaudxmbuYIs0=
AQ7vwjrXlci79BrblbiILS1cxCeoRlcQuXXUy-EinLuqLkW$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 19 2026 - 06:32:41 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST