Hi Casey,
In back at the bottom is the safety box which should have a breaker style
power switch on the left, a few small "bricks" going across, and ends with
the accessory outlets on the right. Check that the switch is up/on first
then the more narrow piece just right of the switch should be a fuse holder
with a standard glass mini fuse in it (I think there's one fuse, possibly
2). Check to see if that's blown.
Good luck,
Andrew
On Mon, May 18, 2026 at 8:27 PM Casey Simons via groups.io wrote:
> Hey all, we just had a power outage which flipped everything off. New to
> me is that our advance III consoles main power switch will not stay
> depressed. My guess is a blown fuse somewhere in the console. Does anyone
> have experience dealing with this?
>
> C
>
> Casey Simons
> NMR Director
> CAMCOR Nanofabrication and Imaging Facility
> University of Oregon
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.