Re: [AMMRL] Executing "SU" never completes VNMRJ 4.2 VNMRS. #agilent #Troubleshooting #varian #VarianAgilent

From: <brcherry_at_asu.edu>
Date: Wed, 22 May 2024 15:48:16 -0700

I checked all the connectors nothing tarnished.  I blew off any dust. 
reconnected.   no change. REF GEN has a stable 80 MHz adn 10 MHz outputs
that are teh same as the other VNMRS consoles that we have.  I then swapped out
the LOCK XCVR board with the one from the MR400.  Just in case.  as before, setacq
runs just fine.  but SU hangs indefinitely, with no errors or messages.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1277): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1277__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwqG3RGWYru1P=
Vkj1ksyoYJFGupALywZXtBUjebJ9ig9qTPLb_1uFsIWTg0pmeN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106247663=
/7559972__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwqG3RGWYru1PVkj1ksyoYJFGupA=
LywZXtBUjebJ9ig9qTPLb_1uFsIWVRu-SUP$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwqG3RGWYru1PVkj1ks=
yoYJFGupALywZXtBUjebJ9ig9qTPLb_1uFsIWeJ5CJLc$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwq=
G3RGWYru1PVkj1ksyoYJFGupALywZXtBUjebJ9ig9qTPLb_1uFsIWXtZ7RgN$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwqG3RGWYru1PVkj1ksyo=
YJFGupALywZXtBUjebJ9ig9qTPLb_1uFsIWaGsOTS8$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!QyCMUcAfSCc4DcNOTNwqG3RG=
WYru1PVkj1ksyoYJFGupALywZXtBUjebJ9ig9qTPLb_1uFsIWfKqMUPE$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed May 22 2024 - 15:48:20 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST