Re: [AMMRL] Executing "SU" never completes VNMRJ 4.2 VNMRS. #agilent #Troubleshooting #varian #VarianAgilent

From: <brcherry_at_asu.edu>
Date: Wed, 22 May 2024 16:09:44 -0700

No answers...but I was able to get the master controller to display what it
did and where it hangs with I do and SU.

rlogin master1
-> ffdir
-> DebugLevel=1

then I did and su.   this is the result:

MR400.MRRC.com:vnmr1 124>rlogin master1

-> ffdir
boot.ini 1384 bytes
rfexec.o 694826 bytes
vxWorks405gpr.bdx 1320720 bytes
nddslib.o 5844541 bytes
rf_amrs.4th 35731 bytes
ddrexec.o 697047 bytes
gradientexec.o 502689 bytes
nvScript 1025 bytes
nvlib.o 776297 bytes
masterexec.o 1022970 bytes
pfgexec.o 475679 bytes
lockexec.o 240954 bytes

12 file(s) 11613863 bytes
3983705 bytes free
FFS size: 15597568
value = 0 = 0x0
-> DebugLevel=1
DebugLevel = 0x1860d5c: value = 1 = 0x1
-> 23:01:32  'nvacq/42x/downLink.c' line: 313,  --------  C_DNWLD_START Received, give Semaphore, timeout: 9
23:01:32  'nvacq/42x/downLink.c' line: 440, Receiving: 'exp1lc0', total size: 440, size sent:440, offset: 0
23:01:32  'nvacq/42x/downLink.c' line: 405, Recv'd: Acodes: 0, Patterns: 0, Tables: 0
23:01:32  'nvacq/42x/AParser.c' line: 527, APstart()-  ExpName: 'exp1'  Num Acodes: 1 Num Tables: 0, startFid: 0
23:01:32  'nvacq/42x/nexus.c' line: 381, Cntlr: 'lock1' Acknowledged ROLLCALL, with: 'Ready & Idle'.
23:01:32  'nvacq/42x/nexus.c' line: 382, Cntlr: 'lock1', ConfigInfo: 0
23:01:32  'nvacq/42x/nexus.c' line: 381, Cntlr: 'pfg1' Acknowledged ROLLCALL, with: 'Ready & Idle'.
23:01:32  'nvacq/42x/nexus.c' line: 382, Cntlr: 'pfg1', ConfigInfo: 0
23:01:32  'nvacq/42x/nexus.c' line: 381, Cntlr: 'ddr1' Acknowledged ROLLCALL, with: 'Ready & Idle'.
23:01:32  'nvacq/42x/nexus.c' line: 382, Cntlr: 'ddr1', ConfigInfo: 0
23:01:32  'nvacq/42x/nexus.c' line: 381, Cntlr: 'rf2' Acknowledged ROLLCALL, with: 'Ready & Idle'.
23:01:32  'nvacq/42x/nexus.c' line: 382, Cntlr: 'rf2', ConfigInfo: 0
23:01:32  'nvacq/42x/nexus.c' line: 381, Cntlr: 'rf1' Acknowledged ROLLCALL, with: 'Ready & Idle'.
23:01:32  'nvacq/42x/nexus.c' line: 382, Cntlr: 'rf1', ConfigInfo: 0
23:01:32  'nvacq/42x/downLink.c' line: 440, Receiving: 'exp1f0', total size: 320, size sent:320, offset: 0
23:01:32  'nvacq/42x/downLink.c' line: 405, Recv'd: Acodes: 1, Patterns: 0, Tables: 0
23:01:32  'nvacq/42x/AParser.c' line: 537, APstart(): Waiting for at least 1 codes to be received
23:01:32  'nvacq/42x/AParser.c' line: 554, CodesDwnLded: 1, freeBufs: 198, total Acodes: 1
23:01:32  'nvacq/42x/downLink.c' line: 661, nDownldCodes: 1, nDownldTables: 0, nDownldPatterns: 0
23:01:32  'nvacq/42x/AParser.c' line: 781, getAcodeSet: Find acode buffer: exp1f0
23:01:32  'nvacq/42x/A32Interp.c' line: 1610, A32Interp(): pAcodeId->cur_acode_size: 296 bytes
23:01:32  'nvacq/42x/A32Interp.c' line: 1779, SENDSYNC: post delay 320 ticks, prepScan: 0
23:01:32  'nvacq/42x/shandler.c' line: 402, addPendingSWItr: add Acode 0xacd00207 to pending list
23:01:32  'nvacq/42x/A32Interp.c' line: 5459, ROLLCALL
23:01:32  'nvacq/42x/A32Interp.c' line: 5463, ROLLCALL: Controllers used in PS: 'rf1 rf2 ddr1 pfg1 '
23:01:32  'nvacq/42x/downLink.c' line: 345,  ---------- C_DNWLD_CMPLT Recieved, give Semaphore
23:01:32  'nvacq/42x/A32Interp.c' line: 5502, ROLLCALL: All required Controllers present.
23:01:32  'nvacq/42x/shimHandlers.c' line: 583, establishShimType: shimset 4
23:01:32  'nvacq/42x/shimHandlers.c' line: 637, establishShimType: shimType 9
23:01:32  'nvacq/42x/A32Interp.c' line: 5741, z0 = 0xfffff405
23:01:32  'nvacq/42x/A32Interp.c' line: 5749, lockgain = 0x26
23:01:32  'nvacq/42x/A32Interp.c' line: 5753, lockpower = 0x23
23:01:32  'nvacq/42x/A32Interp.c' line: 5774, lockphase = 0xab
23:01:32  'nvacq/42x/A32Interp.c' line: 5780, lockfreq = 7.598600 0x401e64f7 0x65fd8ad8
23:01:32  'nvacq/42x/A32Interp.c' line: 5786, lock mode word = 0xa5030003
23:01:32  'nvacq/42x/A32Interp.c' line: 5823, TINFO 0=1
23:01:32  'nvacq/42x/A32Interp.c' line: 5823, TINFO 1=1
23:01:32  'nvacq/42x/A32Interp.c' line: 5557, PNEUTEST pin=15
23:01:32  'nvacq/42x/A32Interp.c' line: 5560, pneuErr=0, pneuLock=15
23:01:32  'nvacq/42x/A32BridgeFuncs.c' line: 908, SETVT: type=2 pid=440 temp=30 tmpoff=200 intlk=0
23:01:32  'nvacq/42x/shandler.c' line: 402, addPendingSWItr: add Acode 0xacd001fa to pending list
23:01:32  'nvacq/42x/shandler.c' line: 402, addPendingSWItr: add Acode 0xacd00207 to pending list
23:01:32  'nvacq/42x/A32BridgeFuncs.c' line: 927, ==========>  SetVT: Suspend Parser
23:01:32  'nvacq/42x/cntrlFifoBufObj.c' line: 760, fifoBufForce(): Send Buf: *0x3c24020=0x8f00000, entries: 30
23:01:32  'nvacq/42x/A32BridgeFuncs.c' line: 267, startCntrlFifo():  fifoStarted4Exp: 'NO'
23:01:32  'nvacq/42x/nexus.c' line: 400, Cntlr: 'pfg1', Reported: 'Ready For Sync'.
23:01:32  'nvacq/42x/nexus.c' line: 400, Cntlr: 'ddr1', Reported: 'Ready For Sync'.
23:01:32  'nvacq/42x/nexus.c' line: 400, Cntlr: 'rf2', Reported: 'Ready For Sync'.
23:01:32  'nvacq/42x/nexus.c' line: 400, Cntlr: 'rf1', Reported: 'Ready For Sync'.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1278): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1278__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2fOWlVTr-ui=
Vsck2eliwYMbA1xBcpwaAdYRcr6Fs0YNZfJTNedESxVssYS25K$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106247663=
/7559972__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2fOWlVTr-uiVsck2eliwYMbA1x=
BcpwaAdYRcr6Fs0YNZfJTNedESxVigNH4j5$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2fOWlVTr-uiVsck2e=
liwYMbA1xBcpwaAdYRcr6Fs0YNZfJTNedESxVloK67UJ$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2=
fOWlVTr-uiVsck2eliwYMbA1xBcpwaAdYRcr6Fs0YNZfJTNedESxVnJQuCWE$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2fOWlVTr-uiVsck2eli=
wYMbA1xBcpwaAdYRcr6Fs0YNZfJTNedESxVmxKbmTr$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!QTSCuInKHBur1MqUwZl2fOWl=
VTr-uiVsck2eliwYMbA1xBcpwaAdYRcr6Fs0YNZfJTNedESxVok4gCpH$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed May 22 2024 - 16:09:47 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST