Re: [AMMRL] Bruker Spinners #Bruker #Hardware

From: Joseph Dumais <josephdumais_at_boisestate.edu>
Date: Tue, 14 May 2024 14:37:56 -0600

Hi Samrat,

Topspin has three depictions of the status of the spinner/sample in the
status bar with respect to the spinner/sample position. One is the with a
question mark, one is the spinner/sample with a profile of a hand picking
it up from the top of the stack and the other is the spinner/sample loaded
into the coil. We now have SampleCases on both systems and I am assuming
that the hand picking it up would be momentarily displayed meaning that the
detector at the upper end of the barrel detects the rotor/sample but I have
never watched that symbol since ours has always worked.

Which depiction do you see when it refused to load a spinner?
Does the sample changer time out and return the rotor or does it park there
with the air on?
My understanding of how the rotor's are detected in the lower end of the
barrel is that Bruker momentarily rotates the sample to detect the light to
dark transition. I think they use the sample optical detection at the top
of the barrel but I don't know if that location needs the light to dark
transition for detection?

Regards,

Joe
*******************************
Joe Dumais, Ph.D.
Associate Research Professor/Nuclear Magnetic Resonance Spectroscopist
Boise State University
College of Arts and Sciences
Department of Chemistry and Biochemistry
1910 University Drive, Boise, Idaho 83725-1520
(208) 426 4913
http://chemistry.boisestate.edu/
*******************************


On Tue, May 14, 2024 at 2:17 PM Samrat Amin via groups.io wrote:

> Hi All,
>
> We are having some issues with Bruker spinners being recognized/loaded
> into magnets. If anyone has any insight into how samples are recognized
> in Bruker's sample changers, we would be grateful to hear more about it.
> I'll provide some information below on our specific issues:
>
> *24 position temp regulated sample changer (SampleCase): * We have this
> unit installed on our 850, and it clearly favors the clear/KEL-F 3mm
> spinners (which seem a bit heavier than the blue POM spinners). These
> clear spinners usually load without any problems. However, Bruker's 3mm
> blue (POM) spinners will not load into the magnet most times. The changer
> will pickup the sample and float it over the bore (indicating it's not a
> gas pressure issue), but it will not lower it into the magnet. We also
> have some semi-clear 3D printed spinners, and these will not load either
> most of the time.
>
> *60 position sample changer (Sample Xpress):* We have two units
> installed (on 400,500) and both units have the same exact model number. On
> the newer unit (500), we are able to use 3D printed spinners without any
> problems. On the older unit (400), it refuses to load the same exact
> spinners. Bruker's blue POM spinners work fine on both instruments as do
> their other spinners that are quite different in weight (KEL-F, PEEK). We
> have also weight matched the 3D printed spinners, so that doesn't seem to
> be a factor on these units.
>
> We are wondering if its the color, weight, or material that is causing the
> inconsistencies in these automation units (ie, are the sensors in the
> sample changers particular to any of these properties). So far, everything
> seems to indicate that weight is not the problem here (ie, the 60 position
> changers even handle much heavier j. young type tubes).
>
> Thanks,
> Samrat Amin
> Magnetic Resonance Research Center
> Arizona State University
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1251): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1251__;!!PvDODwlR4mBZyAb0!Q6ecRJ7wG1I5EyUtSRpWBeZ0cxHnTU=
5bcmuvPDo43_groYejwIEP9809AA82R3hjdetmdFhHPPqHJqax-vxiheRPnNM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106102202=
/7559972__;!!PvDODwlR4mBZyAb0!Q6ecRJ7wG1I5EyUtSRpWBeZ0cxHnTU5bcmuvPDo43_gro=
YejwIEP9809AA82R3hjdetmdFhHPPqHJqax-vxioumP9Y8$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Q6ecRJ7wG1I5EyUtSRpWBeZ0cxHnTU5bcmuvPD=
o43_groYejwIEP9809AA82R3hjdetmdFhHPPqHJqax-vxiC-Q3-AU$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Q6ecRJ7wG1I5EyUtSRpWBeZ0cxHnTU5bcm=
uvPDo43_groYejwIEP9809AA82R3hjdetmdFhHPPqHJqax-vxiR4vFstU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 14 2024 - 13:38:37 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST