Dear all,

An update on the pulsecal issues that I reported earlier. It turned out that
it was the 1H2D board in the preamp that was causing the issue for pulsecal
and also homodecoupling while other proton pulses were only marginally affected.
We ended up replacing the board to fix the problem.