Hi Stu,
I found a couple issues with Alma and Topspin but now it all works great. For
IPSO based systems like we have (AVIIIs), the tftp process needs to be enabled
or started or the IPSO won’t boot. By default it is not enabled. A
second issue is that the sticky bit on the /tmp directory gets reset on boot
which causes a permission problem with Topshim if you change linux users. One
needs to unset the sticky bit on boot. I put a command in the cron to do so.
I did see your error about CPR listener at one point but I believe a reboot
solved the issue but it might have been on one of my failed attempts which
caused me to start over from scratch. Sorry I can’t help. Finally,
Bruker recommends installing Codemeter 8 from the codemeter website. But one
should install Codemeter 7 from the Topspin 3.7.0 installation package first
and then upgrade to Codemeter 8 runtime. There was a library package that is
installed by the Bruker installer that is necessary.
Hope this helps people.
John
> On May 2, 2024, at 10:20 AM, Parnham, Stu <stitch_at_email.unc.edu> wrote:
>
> Hey Martha,
>
> Many thanks for the reply. I thought that to begin with but after watching
> the dialogue in the shell Topspin seems to see the CodeMeter license or at
> least doesn’t complain about it.
>
> Stu
>
> -------------------
> Stuart Parnham Ph.D.
> Biomolecular NMR
> Marsico Hall, SB205A
> University of North Carolina
> Phone:(919)-966-6184
> From: main_at_ammrl.groups.io Behalf Of Martha Morton via groups.io
> Sent: Thursday, May 2, 2024 9:31 AM
> To: main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Topspin3.7 issues
>
> Stu,
>
> Is it a CodeMeter problem?
>
> Martha Morton
>
> On Thu, May 2, 2024 at 8:01 AM Parnham, Stu via groups.io wrote:
> AMMRL’s
>
> I’ve recently installed ALMA 9.3 onto a new PC and installed Topspin3.7
> using Bruker’s instructions and repositories. Everything went smoothly
> until trying to start Topspin which then gives a ‘CPR listener on the
> following TCP/IP port numbers but all seem to be unavailable’ error message.
> The network appears to be configured correctly.
>
> I Has anyone else run into similar issues?
>
> Kindest regards
>
> Stu
>
> -------------------
> Stuart Parnham Ph.D.
> Biomolecular NMR
> Marsico Hall, SB205A
> University of North Carolina
> Phone:(919)-966-6184
>
> --
> Martha Morton, PhD
> Director of Research Instrumentation, University of Nebraska Lincoln
> Safety Co-Chair, Department of Chemistry
> Nebraska ACS, Chair 2021
> ACS, MWRM Chair 2021
>
John Decatur, Ph. D.
Senior Research Scientist, Director of Chemical Instrumentation
and Precision Biomolecular Characterization Facility
Department of Chemistry
3000 Broadway, Mailcode 3179
Columbia University
New York, NY 10027
212-854-2155 (voice)
212-932-1289 (fax)
jdd13_at_columbia.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1213): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1213__;!!PvDODwlR4mBZyAb0!QgoxMBtZnjfRzS_pjMzP7cMWa97gXd=
wNCB63Np2cZmk2LFBH50vJmI8j_l1ZjpaJu3Kmp_OKGQ_ofKaL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105862048=
/7559972__;!!PvDODwlR4mBZyAb0!QgoxMBtZnjfRzS_pjMzP7cMWa97gXdwNCB63Np2cZmk2L=
FBH50vJmI8j_l1ZjpaJu3Kmp_OKGQ9RYtHE$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu May 02 2024 - 08:36:47 MST