Re: [AMMRL] : Puzzle of glucose HSQC

From: Roberto R. Gil via groups.io <rgil=andrew.cmu.edu_at_groups.io>
Date: Wed, 4 Mar 2026 19:07:10 -0500

Hello Spinlanders,

Those "COSY Artefacts" were present in Bruker HSQc pulse programs until
version "hsqcetgpsisp2.2" and were finally removed in version
"hsqcetgpsisp2.3"

;hsqcetgpsisp2.3
;avance-version (20/03/03)
;HSQC
;2D H-1/X correlation via double inept transfer
; using sensitivity improvement
;phase sensitive using Echo/Antiecho-TPPI gradient selection
;with decoupling during acquisition
;using trim pulses in inept transfer
;using shaped pulses for all 180degree pulses on f2 - channel
;with gradients in back-inept
;with suppression of COSY type artefacts

Cheers
Roberto

On Wed, Mar 4, 2026 at 5:58 PM Gareth Morris via groups.io wrote:

> I think these “COSY artefacts” may be the HSQC analogue of the artefacts
> seen in HETCOR when one of the one-bond 13C satellites lands close to the
> shift of a neighbouring proton, causing strong coupling. These were
> described in "'Virtual Coupling' in Heteronuclear Chemical Shift
> Correlation by Two-Dimensional NMR. A Simple Test". G.A. Morris and K.I.
> Smith, J. Magn. Reson. 65, 506-509 (1985), and "Analysis of 'Virtual
> One-Bond Coupling' Effects in Heteronuclear Chemical Shift Correlation 2D
> NMR Spectra". G.A. Morris and K.I. Smith, Mol. Phys. 61, 467-483, (1987).
>
> Gareth
>
>
>
> On 4 Mar 2026, at 16:02, Walt Massefski via groups.io wrote:
>
> This Message Is From a New External Sender
> You have not previously corresponded with this sender. Please exercise
> caution when opening links or attachments included in this message.
> We generally see COSY artifacts in our HSQCs (at least for strong
> samples), but I would suggest that they are actually a good artifact - they
> are small, so you won’t confuse the COSY signals with real HSQC peaks, and
> when you see them, it gives you good nearest neighbor connections. It’s
> like getting H2BC data for free!
> Walt
>
> -------------------------------
>
> Walter Massefski, Ph.D.
> Director
> Department of Chemistry Instrumentation Facility
> Massachusetts Institute of Technology
> 77 Mass. Ave. Rm: 18-0090
> Cambridge, MA 02139
> T: (617) 253-2016
> wwm_at_mit.edu
>
> On Mar 3, 2026, at 4:37 PM, Anklin, Clemens via groups.io wrote:
>
> Hi Weixing
>
> HSQC's can show COSY type peaks when the acquisition time in the indirect
> dimension gets long. Would these peaks by any chance align with the C4
> protons? The sequence hsqcedetgpsisp2.4 suppresses these COSY type
> artefacts.
>
> Clemens
>
> On 3/3/2026 2:38 PM, Zhang, Weixing via groups.io wrote:
>

> Dear Colleagues,

> The attached picture shows the HSQC cross-peak of C3 and C5 in natura abundance β-glucose.
> Coupling to the C2 and C4 protons makes C3 appear as a triplet, but I am
> unsure what the small peak located ~32 Hz to the right of the main signal
> represents.
> Similarly, for C5 there is also a small peak on the right side,
> approximately 20 Hz away from the main peak.

> I would appreciate any insight into the origin of these additional peaks.

> Thank you in advance.

> Best regards,

> Weixing
>
> ------------------------------
>

--

“The tragedy of life doesn't lie in not reaching your goal. The tragedy
lies in having no goal to reach. It isn't a calamity to die with dreams
unfilled, but it is a calamity not to dream. It is not a disgrace not treach
the stars, but it is a disgrace to have no stars to reach for.”

-Benjamin E. Mays (American educator, Clergyman, 1895-1984)-

Roberto R. Gil, Ph.D.
Professor
Director, Department of Chemistry NMR Center
Editor in-Chief, Magnetic Resonance in Chemistry
Department of Chemistry
Carnegie Mellon University
4400 Fifth Ave
Pittsburgh, PA 15213
412-268-4313
412-268-1061 FAX
rgil_at_andrew.cmu.edu
Public Scientific Profile at https://orcid.org/0000-0002-8810-5047

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3267): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3267__;!!PvDODwlR4mBZyAb0!SZq-zW3VGiXhi3pPwvhVRgPAJUF6m3=
9gw0qcLLlGuVUSgv3ZSxNJeqEEu1Kwv9pAoQ76i0tANmg2o7c2aG5vrioSPFIM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118120016=
/7559972__;!!PvDODwlR4mBZyAb0!SZq-zW3VGiXhi3pPwvhVRgPAJUF6m39gw0qcLLlGuVUSg=
v3ZSxNJeqEEu1Kwv9pAoQ76i0tANmg2o7c2aG5vrsTHeV3I$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 04 2026 - 16:31:18 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST