I think these “COSY artefacts” may be the HSQC analogue of
the artefacts seen in HETCOR when one of the one-bond 13C satellites lands
close to the shift of a neighbouring proton, causing strong coupling. These
were described in "'Virtual Coupling' in Heteronuclear Chemical Shift
Correlation by Two-Dimensional NMR. A Simple Test". G.A. Morris and K.I.
Smith, J. Magn. Reson. 65, 506-509 (1985), and "Analysis of 'Virtual
One-Bond Coupling' Effects in Heteronuclear Chemical Shift Correlation
2D NMR Spectra". G.A. Morris and K.I. Smith, Mol. Phys. 61, 467-483, (1987).
Gareth
On 4 Mar 2026, at 16:02, Walt Massefski via groups.io wrote:
We generally see COSY artifacts in our HSQCs (at least for strong samples),
but I would suggest that they are actually a good artifact - they are small,
so you won’t confuse the COSY signals with real HSQC peaks, and when
you see them, it gives you good nearest neighbor connections. It’s
like getting H2BC data for free!
Walt
-------------------------------
Walter Massefski, Ph.D.
Director
Department of Chemistry Instrumentation Facility
Massachusetts Institute of Technology
77 Mass. Ave. Rm: 18-0090
Cambridge, MA 02139
T: (617) 253-2016
wwm_at_mit.edu<mailto:wwm_at_mit.edu>
On Mar 3, 2026, at 4:37 PM, Anklin, Clemens via groups.io wrote:
Hi Weixing
HSQC's can show COSY type peaks when the acquisition time in the indirect
dimension gets long. Would these peaks by any chance align with the C4
protons? The sequence hsqcedetgpsisp2.4 suppresses these COSY type artefacts.
Clemens
On 3/3/2026 2:38 PM, Zhang, Weixing via groups.io wrote:
Dear Colleagues,
The attached picture shows the HSQC cross-peak of C3 and C5 in natural abundance β-glucose.
Coupling to the C2 and C4 protons makes C3 appear as a triplet, but I am unsure what
the small peak located ~32 Hz to the right of the main signal represents.
Similarly, for C5 there is also a small peak on the right side, approximately 20 Hz
away from the main peak.
I would appreciate any insight into the origin of these additional peaks.
Thank you in advance.
Best regards,
Weixing
________________________________
==========================
====================
Prof Gareth A Morris FRS
Department of Chemistry, University of Manchester
Oxford Road, Manchester M13 9PL, UK
Tel (0) 161 275 4665
g.a.morris_at_manchester.ac.uk
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3266): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3266__;!!PvDODwlR4mBZyAb0!UEycqIGJhPDR5oYnZqXIrdqEz2xV30=
DHMypqxAOVX68m5NI66V1MfZpRqCyEUrxvV-BRtujAa1uGe45IQd6uZ-I6sTAf$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118120016=
/7559972__;!!PvDODwlR4mBZyAb0!UEycqIGJhPDR5oYnZqXIrdqEz2xV30DHMypqxAOVX68m5=
NI66V1MfZpRqCyEUrxvV-BRtujAa1uGe45IQd6uZ3_TFo0Y$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Mar 04 2026 - 14:58:18 MST