From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Tue, 25 Mar 2025 19:44:15 +0000 (UTC)

That's a weird one.  Are there multiple users that it's failing for or
is it just the one that can't get it to work?  Are all the numbers in
the dataset correct--are there pulse lengths and pulse powers set, for example?

One thing I might check would be to see if the user is a member of either the
nmr users or the nmrsu groups.  Maybe there's some weird permission
thing going on where if that user isn't a part of one of those groups they
don't have permission to write to the topspin data directory?  Do you
see any FID being acquired or is the display always blank?  Can you see
anything in gs mode?  And can they acquire anything else--13C for
example?  

Cheers,Mike

On Tuesday, March 25, 2025 at 01:25:50 PM MDT, Brian Antalek via groups.io wrote:

I am running TopSpin 3.2 on an AVII 300 MHz system under the nmrsu user. 
Manual operation and iconnmr automation work fine.But, when I switch to a
different user, I cannot obtain 1H signal. It tunes ok to 1H and locks and
shims fine on D2O, but there is no acquired signal. Before I switched
to a different user, I exited TopSpin.  I also successfully ran ii when
I restarted TopSpin in the other user.  

Has anyone ever seen this? 

Does anyone have a remedy? 

thanks

Brian 

Brian Antalek

Bausch + Lomb, Inc.
1400 N. Goodman Street
Rochester, NY 14609

http://www.bausch.com

brian.antalek_at_bausch.com

(585) 752-8452  mobile



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2167): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2167__;!!PvDODwlR4mBZyAb0!RzAauNjEGALAiA5b3YCt0IgzGjS4zc=
YxiOWo-h7uYRms5W1jYxQIwPA1MOTjkYiCW5ZO0p39_5BFva7POfPa6RXBkfKZ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111903971=
/7559972__;!!PvDODwlR4mBZyAb0!RzAauNjEGALAiA5b3YCt0IgzGjS4zcYxiOWo-h7uYRms5=
W1jYxQIwPA1MOTjkYiCW5ZO0p39_5BFva7POfPa6VigSkb3$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Mar 25 2025 - 12:44:54 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST