Re: [AMMRL] Protune Control Boxes

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Fri, 14 Mar 2025 10:42:47 +0100

Hi Shawn,

while at it you should regrease them all, as the grease will be equally old!
Any machine grease should do, as long as it's new ;-)

Let me know if you need help repairing your probe…

Cheers
Bert

> Am 11.03.2025 um 19:11 schrieb Shawn Wagner via groups.io:
>
> Thanks, Todd, for reposting the instructions, Eugenio, and those that responded privately.
> This was very helpful. I pulled one of the bad one apart and it was quite sticky.
> I plan to use engine assembly grease. I think this should work well on metal parts.
> Shawn
>
> From: main_at_ammrl.groups.io on behalf of Shawn Wagner via groups.io
> Sent: Monday, March 10, 2025 9:46 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Protune Control Boxes
>
>
> Hi Varian Users,
>
> Has anyone else had the issue that the ProTune motors are exerting too much
> force on the tune rods? The probe capacitors for the X channel are being=
> over tuned during the indexing and one broke the gear free from the rod. Unlike
> the highband channel capacitors which have turn limiters to protect the
> capacitors, the X band does not. Is a way to reduce the power to the motors?
>
> Shawn
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifirs (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2151): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2151__;!!PvDODwlR4mBZyAb0!TuFz-S3E6-cSMykbrfRhcp4ycwSNKp=
31yjVgPrCajkIjVCxe73cNOvCfuIpb9IZJmcjoC3A2oiWbYLFXNqTobQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111623088=
/7559972__;!!PvDODwlR4mBZyAb0!TuFz-S3E6-cSMykbrfRhcp4ycwSNKp31yjVgPrCajkIjV=
Cxe73cNOvCfuIpb9IZJmcjoC3A2oiWbYLELlukQCA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Fri Mar 14 2025 - 06:01:26 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST