[AMMRL] Protune Control Boxes

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Mon, 10 Mar 2025 16:46:41 +0000

Hi Varian Users,

Has anyone else had the issue that the ProTune motors are exerting too much
force on the tune rods? The probe capacitors for the X channel are being over
tuned during the indexing and one broke the gear free from the rod. Unlike the
highband channel capacitors which have turn limiters to protect the capacitors,
the X band does not. Is a way to reduce the power to the motors?

Shawn


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.