Re: [AMMRL] 1H-19F HOESY setup tips?

From: brcherry via groups.io <brcherry=asu.edu_at_groups.io>
Date: Mon, 03 Mar 2025 10:13:02 -0800

HI,

I was able to get these experiments to work on our AVIII system and a BBFO
probe.  As Dan mentioned it helps to have a good RF filter.   
I simply used a 19F low pass filter (Z13331 400 0-31P,19F-LP (1H)) on the BB
channel and nothing external on the 1H channel.

I used this paper as guide.  the molecule can be purchased inexpensively
to if you need a reference compound.

https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/full/10.1002/mrc.4904

Brian


-=-=-=-=-=-=-=-=-=-=-=-