Re: [AMMRL] #agilent #Hardware #VarianAgilent

From: Gary Thompson via groups.io <G.S.Thompson=kent.ac.uk_at_groups.io>
Date: Mon, 3 Mar 2025 14:04:24 +0000

I have memories that the capacitors in the PTS can go bad an cause errors
like this, maybe Bert can comment

regards
Gary

Dr Gary S Thompson NMR Facility Manager
CCPN CoI & Working Group Member
Wellcome Trust Biomolecular NMR Facility
School of Biosciences, Division of Natural Sciences
University of Kent, Canterbury, Kent, England, CT2 7NZ

☎:01227 82 7117
✉️: g.s.thompson_at_kent.ac.uk
orchid: orcid.org/0000-0001-9399-7636

On 28 Feb 2025, at 21:15, johnhahn cazent.com <johnhahn_at_cazent.com> wrote:

Hi Penghui,

This would most likely be a problem with the 80 MHz harmonic generator. What
you are seeing, is a phase stability issue with the higher sideband generated
for the highband observe. The fact that it moved with the swap, the channel
board is probably OK. Easy fix.....?! it depends on the problem at hand. We
don't have schematics for these things.......

<Outlook-z1tuourg.png>
​Regards,
​John E. Hahn

CAZ Enterprises, LLC
6320 Fly Rd. Ste 207
East Syracuse, NY 13057
Ph. 315-727-8192
http://www.Cazent.com
________________________________

> From: main_at_ammrl.groups.io on behalf of Lin, Penghui via groups.io
> Sent: Friday, February 28, 2025 14:33
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] #agilent #Hardware #VarianAgilent

Dear AMMRLers,

We had a 600 MHz Agilent DDR2 system.

I recently noticed the proton channel (1st channel) tune has some problems.
It shows very noisy signals (first picture). But the carbon channel (channel
2) is normal.

If I swapped the cables on the VFS between channel 1 and channel 2, and the
proton went back to normal and I am having troubles with carbon tune now.
(proton tune back to normal 2nd picture).

When I check the signals coming out from the VFS on channel 1 and channel 2
with an oscilloscope, the "problematic" channel 1 signals look like 3rd picture
and the normal channel 2 signals look like the last picture.

Does this mean this VFS board has problems? If so, is there an easy way to
fix it?

Thanks,
Penghui






-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2095): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2095__;!!PvDODwlR4mBZyAb0!VPo98x4fWuM4y6uSEfVwNSAVnAGjTY=
XtLs4JxkwyICsvW58T6tI-BgnkMZbM5Gr45kZH8JPbcxvDtpeJEZE17gEfoj-F$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111440908=
/7559972__;!!PvDODwlR4mBZyAb0!VPo98x4fWuM4y6uSEfVwNSAVnAGjTYXtLs4JxkwyICsvW=
58T6tI-BgnkMZbM5Gr45kZH8JPbcxvDtpeJEZE17vzN85Dl$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!VPo98x4fWuM4y6uSEfVwNSAV=
nAGjTYXtLs4JxkwyICsvW58T6tI-BgnkMZbM5Gr45kZH8JPbcxvDtpeJEZE17mZh_ae3$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!VPo98x4fWuM4y6uSEfVwNSAVnAGjTYXtLs4J=
xkwyICsvW58T6tI-BgnkMZbM5Gr45kZH8JPbcxvDtpeJEZE17scchJbB$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!VPo98x4fWuM4y6uSEfVwNSAVnAGjTYXtLs=
4JxkwyICsvW58T6tI-BgnkMZbM5Gr45kZH8JPbcxvDtpeJEZE17qCSpucF$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Mar 03 2025 - 06:04:32 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST