Re: [AMMRL] Topshim question - both out of necessity and curiosity...

From: Frank Holger Foersterling via groups.io <holger=uwm.edu_at_groups.io>
Date: Tue, 3 Jun 2025 14:17:14 -0500

Rajan,

I have only Topspin up to 4.4 on my spectrometers, so I can not speak
for 4.5, but the tuneb option was removed at some point from Topspin 3.7
forward. The file <topspin/conf/instr/topshim/parameters/tuneb file
still exists, but is empty with just a line ##$COMMENT= <DEPRECATED>in it.

So topshim tuneb should skip transverse shimming alltogether, but not
give an error message. You either need to use tunea (transverse shimming
after longitudinal) or write a macro  that contains two topshim commands:

topshim tune tunea

topshim

That should executa a transverse shimming first without any topshim
(that is what the keyword "tune" does), then the regular topshim will be
run You can add any other options to the commands of course.

I have actually not double checked whether spinning is still stopped
automatically for the topshim tune in my latest Topspin versions, if
not, one needs to add commands to sttop and start spinning to the macro.

Not sure thought why Bruker got rid of the tuneb option in the first place.

Holger


On 6/3/25 1:48 PM, DrSpin via groups.io wrote:

> Dear All
>    I am comparing the following scenario between two systems, one of
> them an Avance-II driven by Topspin 3.2 and the other a NEO driven by
> Topspin 4.5.0
> Here are the steps :
>
> * I rotate the sample (20 Hz).
> * I run from TS command line : /topshim tuneb/
> * Watch what happens.
>
> Forgive me if I am adding the obvious, but just in case : the 'tuneb'
> option is supposed to do a simple grid search optimization of these
> transverse shims : X,Y,XZ and YZ.  The 'tuneb' stands for 'before'. 
> Then Topshim optimizes the axial shims.
> AV-II with TS3.2 :    The rotation is stopped,  transverse shims are
> optimized, rotation is turned ON and then,  Topshim of Axial i.e.
> Z-shims happens.   This is the ideal behavior.
> NEO with TS4.5.0 :   As seen from the BSMS status, the rotation is NOT
> stopped,  but the process attempts the transverse shimming (not of any
> use) and then goes onto complete Axial shimming.   Sure enough, when I
> look at 'topshim report', I do not see anything of consequence
> reported about the transverse shimming.  By the way,  the incantation
> /topshim tunea /does not produce anything different. Here /tunea
> /stands for 'after' so that Axial shimming is performed first and then
> transverse shimming is attempted.
> Is this what you observe in your systems ?  This has considerable
> practical implication, especially when we try to run the system via
> IconNMR automation, and when we want to get the best resolution by
> rotation, while avoiding invasive sidebands.
> Thank you for sharing your story and wisdom.
> Best Regards
> Rajan
>

--
Frank H. Foersterling
Instrumentation Specialist
Chemistry 362
Department of Chemistry
University of Wisconsin-Milwaukee
414-229-5035
holger_at_uwm.edu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2372): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2372__;!!PvDODwlR4mBZyAb0!Ql3S9jAAiqqmZPgjSq_ARrdHyke3TH=
y8VEkDtHs-AKYRTW67QvqhKWk7WYf8PmHL_QIZ-FW6QKOaAoOlJQyBw5h95Q$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113453101=
/7559972__;!!PvDODwlR4mBZyAb0!Ql3S9jAAiqqmZPgjSq_ARrdHyke3THy8VEkDtHs-AKYRT=
W67QvqhKWk7WYf8PmHL_QIZ-FW6QKOaAoOlJQzd9QQvIQ$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!Ql3S9jAAiqqmZPgjS=

Received on Tue Jun 03 2025 - 12:48:38 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST